Thank you for purchasing the Braun Pulse oximeter, YK-81C.
This pulse oximeter is a high-quality product tested in
accordance with international ISO safety and performance
standards. This device uses red and infrared light at set
wavelengths along with a photodetector to measure oxygen
saturation in blood and the associated pulse rate.
The Braun Pulse oximeter is intended to be used by consumers in
a home-use setting (such as home setting or places of business).
Please read these instructions carefully before using this
product and keep the instructions and the pulse oximeter in a
safe place.
Indications for use
The Braun Pulse oximeter (YK-81C) is a non-sterile, reusable
device intended for spot checking or intermittent
determination of functional oxygen saturation in arterial blood,
as well as the associated pulse rate by using the tip of the nger
as the measurement site. The device is only intended for adults,
in a home-use environment (such as a home setting or places
of business).
The Pulse oximeter displays the %SpO2, pulse rate, as well as
the pulse amplitude using a vertical bar graph.
This device is factory calibrated and no further calibration is
required by the user.
WARNINGS AND PRECAUTIONS
Please read the manual carefully before use.
Device isn’t suitable for continuous monitoring.
Pulse oximeters are sensitive to motion. Keep hands still while
taking a reading.
Pulse oximeters require sucient blood ow to
obtain proper readings. Poor blood circulation can result in
inaccurate readings. If your hands are cold or if you have poor
circulation, warm your hands by rubbing them together or use
another warming method before attempting to obtain a
reading. A tourniquet, blood pressure cu, other devices or
clothing that cause blood ow restrictions may also result in
inaccurate readings.
Fingernail polish or acrylic nails obstruct the light transmission
through the ngertip and may also result in inaccurate
readings.
Your nger must be clean for a proper reading.
The Pulse oximeter should be clean in order to get a proper
reading. Please ensure the nger clip area is cleaned, following
the cleaning instruction in the manual.
If a reading is dicult to obtain, switch to another nger or to
the other hand. The index or middle nger is recommended for
taking measurements.
There are a number of other conditions which may lead to an
inaccurate reading; including but not limited to, recent medical
tests that included an injection of dyes, use of arterial catheters,
a weak pulse, low levels of hemoglobin in the blood, low
perfusion (the quality of your pulse), elevated levels of
dysfunctional hemoglobin, the strength and type of light that
you are in while using the pulse oximeter, and the existence of
cell phones, radios, and xed transmitters within certain ranges
of the pulse oximeter during use.
This device does not provide a low SpO2 alarm.
DO NOT use on the same nger continuously for over 2 hours.
The device is not appropriate for the constant monitoring of
the user. Prolonged and continuous monitoring may increase
the potential of unexpected changes of dermal conditions
such as putrescence, etc. A feeling of discomfort or pain may
occur if using the device continuously, especially for those
users who experience poor perfusion or immature
dermographia by light collimation.
In addition to the items described in the Warnings and
precautions section, inaccurate measurements may be caused
by factors including but not limited to:
Immersing the device in liquid.
Signicant levels of dysfunctional hemoglobin (such as
carboxyhemoglobin or methemoglobin).
Intravascular dyes such as indocyanine green or methylene
blue.
Measurements may be adversely aected in the presence
of high ambient light such as direct sunlight. In bright light
conditions, cover the sensor area if necessary.
Excessive user movement.
High-frequency electrosurgical interference and debrillators.
Venous pulsations.
Hypotension, severe vasoconstriction, severe anemia, or
hypothermia.
DO NOT modify this equipment without the authorization of
the manufacturer.
English
3


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun YK-81C - Pulse oximeter 1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun YK-81C - Pulse oximeter 1 in the language / languages: All languages as an attachment in your email.

The manual is 0.72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info