Problem/
question
Possible cause/remedy
The atomiser
produces no
or too little
aerosol.
4. Unsuitable medication fluid added
for nebulisation (e.g. too viscous).
The medication fluid should be pre
-
scribed by the doctor.
The output is
too low.
Kinked hose, clogged filter, too much
inhalation solution.
What med
-
ications are
suitable for
inhaling?
Please consult your doctor in rela
-
tion to this matter.
As a rule, all medication that is
suitable and approved for device
inhalation can be inhaled.
There is inha
-
lation solution
residue in the
atomiser.
This is normal and is due to technical
reasons. Stop inhalation once the
atomiser starts to make a notably
different sound.
What should
be taken into
account when
using the
device with
infants and
children?
1. On infants and children, the mask
should cover the mouth and nose
to ensure effective inhalation.
2. On children, the mask should also
cover both the nose and mouth.
It is not a good idea to carry out
nebulisation on someone who
is sleeping, as in this case not
enough of the medication will
reach the lungs.
Note: Inhalation should only be car
-
ried out under the supervision of an
adult and with their assistance and
the child should not be left alone.
Inhalation us
-
ing the mask
takes longer.
This is due to technical reasons. You
breathe less medication per breath
through the holes of the mask than
via the mouthpiece. The aerosol is
mixed with ambient air via the holes.
Problem/
question
Possible cause/remedy
Why should
the atomiser
be replaced
regularly?
There are two reasons for this:
1. To ensure a therapeutically effec
-
tive particle spectrum, the nozzle
hole must not exceed a specific
diameter. Due to mechanical and
thermal stresses, the plastic is
subject to a certain degree of
wear. The atomiser insert [9] is
particularly sensitive. This can also
alter the composition of the aero
-
sol droplets, which directly affects
the effectiveness of the treatment.
2. Regularly changing the atomiser
is also recommended for hygiene
reasons.
Should each
person have
their own
atomiser?
Yes, this is absolutely essential for
hygiene reasons.
11. Technical specifications
Type
IH58
Dimensions
(W x H x D) 110 x 62 x 47 mm
Weight
176 g
Operating
pressure Approx. 0.25 to 0.5 bar
Atomiser filling
volume
Min. 2ml
Max. 6ml
Medication
flowrate Approx. 0.25ml/min.
Sound pressure Max. 45 dBA
(acc. to DIN EN 13544-1 section 26)
Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A
Output: 5 V
; 2 A
Expected
servicelife
400 h
Operating
conditions
Temperature: +10 °C to +40 °C
Relative humidity: 10% to 95%
Ambient pressure: 700 to 1060 hPa
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer IH 58 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer IH 58 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 8,4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info