22 Beurer GL44 mmol/L
Display Blood glucose Action
Very high blood glucose level
Above 35.0 mmol/L
Take another measurement using a new
test strip. If the result is identical:
seek medical attention immediately.
5.6 Functional check with control solution
The control solution is used to test the entire blood glucose monitoring system. This helps to de-
termine whether the monitor and the test strips are working optimally together and whether the test
was performed correctly.
Perform the control solution test if you suspect that the blood glucose monitor and/or the test strips
could be faulty or if you have repeatedly measured unusual blood glucose values. Also test the blood
glucose monitor if it has fallen down or is damaged. The control solution is available separately. Please
observe the additional notes in the Instructions for Use of the control solution for the test.
Important
Do not use third-party control solution. Correct functioning of your measuring device can only be
tested using Beurer control solutions (LEVEL3 + LEVEL4).
Control solution measurements: When using the device, specialist personnel must follow statutory
guidelines.
Do not apply any blood samples or control solutions to the test strips before placing these in the
measuring device.
Performing a functional test using control solution
Warning
To obtain correct results, the measuring device, the test strip and the control solution must be the
same temperature. For the “Functional test using the control solution”, the temperature is to be be-
tween 20 °C and 26 °C.
1
Hold the blood glucose monitor so that the display is facing you.
2
Insert a test strip into the slot on the blood glucose monitor with the contacts first. Make sure
that the front of the test strip is facing you (see “Test strip” on page 12).
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 44 mmol L at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 44 mmol L in the language / languages: English as an attachment in your email.

The manual is 6,14 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 44 mmol L

Beurer GL 44 mmol L User Manual - German - 40 pages

Beurer GL 44 mmol L User Manual - Dutch - 40 pages

Beurer GL 44 mmol L User Manual - French - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info