18 Beurer GL42 mg/dL
Important
Do not use third-party control solution. Correct functioning of the measuring device can only be
tested using the Beurer control solutions (LEVEL5 + LEVEL6).
Control solution measurements: When using the device, specialist personnel must follow statutory
guidelines.
Do not apply any blood samples or control solutions to the test strip before inserting it in the moni-
tor.
Performing a functional test using control solution
Warning
To obtain correct results, the measuring device, the test strip and the control solution must be the
same temperature. For the “Functional test using the control solution”, the temperature is to be be-
tween 20 °C and 26 °C.
1
Hold the blood glucose monitor so that the display is facing you.
2
Insert a test strip into the slot on the blood glucose monitor with the contacts first. Make sure
that the front of the test strip is facing you (see “Test strips” on page 12).
3
The device automatically switches on and briefly shows the initial display. The device is ready
for measurement as soon as the hand
and the symbol flash.
IMPORTANT: Control solutions and blood react to temperature influences in dierent ways.
It is therefore of vital importance that control solution measurement is always performed in
control solution mode. If this mode is not used, results may be obtained that are outside the
target range.
Note
Press the rocker switch “+” or “–” [2] to change to the control mode. “
” is shown on the
display. This means that the result value is not stored in the memory, therefore not influencing
your measured value statistics. Press “+” or “–” [2] again. “ ” disappears from the display
and the value is stored normally in the memory.
4
A clean surface is required to correctly perform a functional test.
Shake the control solution well before use.
Undo the cap and press two drops next to each other on the
clean surface without touching them. Use the second drop for
the measurement.
Note
Do not apply the drop directly to the test strip to avoid contaminating the remaining control
solution in the bottle by touching the test strip with the tip of the bottle.
18


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 42 mg-dl at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 42 mg-dl in the language / languages: English as an attachment in your email.

The manual is 4,58 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 42 mg-dl

Beurer GL 42 mg-dl User Manual - German - 36 pages

Beurer GL 42 mg-dl User Manual - Dutch - 32 pages

Beurer GL 42 mg-dl User Manual - French - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info