Beurer GL34 mg/dL
22
Do not use the control solution if either of these dates has expired (
/ ).
Allow the measuring device, test strips and control solution to reach room temperature
(+20°C to +25°C).
Performing a function test using the control solution
1
Hold the measuring device so that the display is facing you.
2
Insert a test strip into the slit on the measuring device, contacts first. Ensure
that the front side of the test strip is facing towards you (see "3.3 Test strip",
page 11).
3
The device switches on automatically and shows the start display, shortly
followed by the checking symbol “
“. You will hear the voice output instruct-
ing you to apply the blood. As soon as the flashing blood droplet symbol 6 is
displayed, the device is ready to perform measurements.
4
Press the ON/OFF/Forward button to switch to control mode. In control
mode, the measured value is not stored and hence does not distort your
measurement statistics.
5
"CTL" appears on the display.
6
Shake the control solution well before use. Unscrew the cap and squeeze out
a drop of solution. Wipe away the first drop and squeeze out an additional
drop.
7
To prevent the remaining control solution in the bottle from becoming con-
taminated through the tip of the bottle in contact with the test strip, do not
apply the drop directly to the test strip. Instead, apply the drop to a clean
surface, and then apply the drop from there to the blood collection slit on the
test strip. The solution is sucked into the slit. Wipe the tip of the bottle using a
clean, dry paper tissue.
8
Once the slit is filled with solution, the device begins the measurement. The
device counts down for 6 seconds, The measurement is then shown on the
display and simultaneously read out audibly.
9
Check whether the result is within the specified results range for the control
solution. This results range is printed on the test strip container.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 34 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 34 in the language / languages: English as an attachment in your email.

The manual is 5,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 34

Beurer GL 34 User Manual - German - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info