470360
17
Zoom out
Zoom in
Previous page
1/76
Next page
17
Deleting saved values
You can delete saved values individually or else you can
delete all of them.
To delete individual values, first select the stored value
you wish to delete and keep memory button M1 or M2
depressed. The display starts blinking. Keep holding the
button depressed until the display has completely cleared.
To delete an entire user memory, you have to select the
appropriate user memory. The average is displayed and
indicated with
. Hold the memory button M1 or M2
depressed until the display starts blinking and keep the but-
ton depressed until the display completely clears.
8. Error recovery
Error messages may appear if
you move or talk during measurement ( appears in the
display),
the cuff tube is not inserted properly ( appears in the
display),
when changing the batteries, the on/off button was pressed,
the cuff is inflated for longer than 20 seconds ( appears
in the display),
blood pressure is unusually high or low ( appears in
the display),
inflation pressure is higher than 300 mmHg ( appears
in the display).
In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do not
move or talk. Put the batteries back in if necessary, or else
replace them.
9. Cleaning
Clean your blood pressure monitor carefully using a slightly
damp cloth only.
Do not use detergents or solvents.
On no account must you immerse the computer in water,
otherwise liquid can enter it and cause demage.
When storing the device, make sure that no heavy objects
are placed on top of it. Remove the batteries. The cuff tube
should not have any sharp kinks.
10. Specifications
Model No. BM 20
Method of
Measurement
Oscillometric, non-invasive blood pressure
measurement on the upper arm
Range of Meas-
urement
Cuff pressure: 0 300 mmHg,
systolic: 50 – 250 mmHg,
diastolic: 40 –180 mmHg,
pulse: 40 –160 beats/minute
Display
accuracy
systolic: ± 3 mmHg /diastolic: ± 3 mmHg /
pulse: ± 5 % of the isplayed value
Measuring
uncertainty
max. permissible standard deviation accord-
ing to clinical testing in accordance with
EN 1060-4: systolic: 8 mmHg /
diastolic: 8 mmHg
Sets of Memory 2 x 30 memory places
Unit Dimension L110 x W170 x H58 mm
Unit Weight Approx. 265 g
Cuff Size
Cuff for upper arm circumference of 22 32 cm
17


Need help? Post your question in this forum.

Forumrules
1

Forum

beurer-bm-20

Reset search

  • Sir switch not working only some number shows stop Submitted on 29-6-2022 at 15:28

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer BM 20 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer BM 20 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 7,46 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer BM 20

Beurer BM 20 User Manual - Dutch - 12 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info