7.3 Resistance ranges
No-load voltage: approx. 1.3 V - 3.3 V, max. test current 2.5 mA.
42 MΩ-range, response time approx. 20 s
Measuring range Resolution Accuracy
Accuracy overload
protection
420 Ω 0,1 Ω ± (1.2 % of reading + 8 digits)
600 V
eff
600
V
DC
4,2 kΩ 1 Ω ± (0.9 % of reading + 4 digits)
600 V
eff
600
V
DC
42 kΩ 10 Ω ± (0.9 % of reading + 4 digits)
600 V
eff
600
V
DC
420 kΩ 100 Ω ± (1.2 % of reading + 4 digits)
600 V
eff
600
V
DC
4,2 MΩ 1 kΩ ± (1.2 % of reading + 4 digits)
600 V
eff
600
V
DC
42 MΩ 10 kΩ ± (3.0 % of reading + 8 digits)
600 V
eff
600
V
DC
7.4 Diode testing
The stated measurement accuracy applies in the range between 0.4 V and
0.8 V.
Overload protection for diode testing: 600 V
eff
/ 600 V
DC
Measuring
range
Resolution Accuracy
max. meas.
current
Max. no-load
voltage
0.1 mV
± (1.5 % of reading + 5 digits)
1,5 mA
3,3 V
7.5 Continuity testing
The integrated buzzer sounds at resistances R < 50 Ω.
7.6 AC ranges
(Multimeter with current transducer attachment. Current transducer grips single
live conductor wire).
Measurement accuracy is ± (% of reading + number of digits) at a temperature
of 23 °C ± 5 °C.
Max. current of enclosed current transducer 300 A!
Measuring range
Resolution Output voltage Accuracy
300 A 0.1 A 1 mV/ 0.1 A
± (1.5 % of reading + 5 digits)
in frequency range 40 Hz - 300 Hz
8. Measuring with the BENNING MM 4
8.1 Preparation for measurement
Store and use the BENNING MM 4 only under the correct temperature condi-
tions specified. Always avoid prolonged exposure to sunlight.
- Check nominal voltage and current data on the black safety test lead and red
test probe. The black test lead and the red test probes supplied correspond
to the BENNING MM 4 in nominal voltage and nominal current.
- Check insulation of the safety test lead and red test probe. If the insulation
is damaged, discard the lead and test probes immediately.
- Check the continuity of the safety test lead. If the conductor in the safety
test lead is interrupted, discard the safety test lead immediately.
- Before selecting another function at the sliding switch 2 or function button
4, the safety test lead and red test probes must first be disconnected from
the measurement point.
- Strong sources of interference in the vicinity of the BENNING MM 4 may
cause unstable or incorrect readings.
8.2 Voltage measurement
Always observe the maximum voltage to earth potential!
Electrical hazard!
The maximum voltage which may be applied to the sockets of the Multimeter with
- COM socket 7, marked black,
- V-Ω socket (positive) 8 for voltage and resistance measurements, continu-
ity and diode testing (marked red) of the BENNING MM 4 with reference to
earth potential is 600 V.
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning MM 4 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning MM 4 in the language / languages: English, German, Dutch, French, Italian, Polish, Swedish, Turkish, Spanish as an attachment in your email.

The manual is 6,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info