577174
17
Zoom out
Zoom in
Previous page
1/97
Next page
09/ 2013
BENNING CM 1-1/ 1-2
11
against solid impurities of a diameter > 2.5 mm
0 – second index: no protection against water,
- Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less than
80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less than
75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less than
45 %,
- Storage temperature: The BENNING CM 1-1/ 1-2 can be stored at tempera-
tures between - 20 °C and + 60 °C (air humidity of 0 to 80 %). During stor-
age, the battery should be removed.
7. Electricalspecications
Note: The measuring accuracy is specied as the sum of:
- a relative part of the measured value and
- a number of digits (i.e. counting Stepps of the last digit).
This measuring accuracy applies for temperatures from 18 °C to 28 °C and a
relative air humidity less than 80 %.
7.1 DC voltage ranges ( BENNING CM 1-2)
The input resistance is 10 MΩ.
Measuring
range
Resolution Measuring accuracy
Overload
protection
200 V 0.1 V ± (1.0 % of the measured value + 2 digits) 600 V
eff
600 V 1 V ± (1.0 % of the measured value + 2 digits) 600 V
eff
7.2 AC voltage ranges ( BENNING CM 1-2)
The input resistance is 10 MΩ in parallel 100 pF.
Measuring
range
Resolution
Measuring accuracy *
1
within the frequency
range 50 Hz - 500 Hz
Overload
protection
200 V 0.1 V ± (1.5 % of the measured value + 5 digits) 600 V
eff
600 V 1 V ± (1.5 % of the measured value + 5 digits) 600 V
eff
7.3 AC current ranges
BENNING CM 1-1
Measuring
range
Resolution
Measuring accuracy *
1
within
the frequency range 50 Hz - 60 Hz
Overload
protection
20 A 0.01 A ± (3.0 % of the measured value + 5 digits) 600 A
eff
200 A 0.1 A ± (2.0 % of the measured value + 5 digits) 600 A
eff
400 A 1 A ± (2.0 % of the measured value + 5 digits) 600 A
eff
BENNING CM 1-2
Measuring
range
Resolution
Measuring accuracy *
1
*
2
within
the frequency range 50 Hz - 60 Hz
Overload
protection
40 A 0.1 A ± (1.9 % of the measured value + 10 digits) 600 A
eff
200 A 0,1 A ± (1.9 % of the measured value + 5 digits) 600 A
eff
400 A 1 A ± (1.9 % of the measured value + 5 digits) 600 A
eff
*
1
The measured value is obtained by mean value rectication and is displayed
as r.m.s. value. Its calibration is adapted to a sinusoidal curve.
*
2
The stated accuracy is specied for conductors that are centrally clamped
by means of the current clamp
L
(see gure 4 AC current measurement).
For conductors that are not centrally clamped, an additional error of 1,5 % of
the displayed value has to be considered.
7.4 Resistance measuring range and acoustic continuity test
( BENNING CM 1-2)
Overload protection: 600 V
eff
Measuring
range
Resolution Measuring accuracy
Max. open-
circuit voltage
200 0.1 ± (1,0 % of the measured value + 5 digits) 1.3 V
2 kΩ 1 ± (0.7 % of the measured value + 2 digits) 1.3 V
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning CM1.1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning CM1.1 in the language / languages: English, German, Dutch, French, Italian, Polish, Swedish, Turkish, Spanish as an attachment in your email.

The manual is 4,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email