577172
8
Zoom out
Zoom in
Previous page
1/39
Next page
2. Safety note
The instrument is built and tested in accordance with
DIN VDE 0411 part 1/ EN 61010-1
DIN VDE 0411-2-032/ EN 61010-2-032
and has left the factory in perfectly safe technical state.
To maintain this state and ensure safe operation of the appliance tester, the user
must observe the notes and warnings given in these instructions at all times.
The unit may be used only in power circuits within the overvoltage
category III with a conductor for 600 V max. to earth .
Remember that work on electrical components of all kinds is
dangerous. Even low voltages of 30 V AC and 60 V DC may be
dangerous to human life.
Before starting the appliance tester up, always check it as well
as all cables and wires for signs of damage.
Should it appear that safe operation of the appliance tester is no longer possible,
it should be shut down immediately and secured to prevent it being switched
on accidentally.
It may be assumed that safe operation is no longer possible:
-
if the instrument or the measuring cables show visible signs of damage, or
- if the appliance tester no longer functions, or
- after long periods of storage under unfavourable conditions, or
- after being subjected to rough transport.
In order to avoid danger,
- do not touch the bare prod tips of the measuring cables
connectors,
- insert the measurement lines in the appropriately
designated measuring sockets on the multimeter
3. Scope of delivery
The scope of delivery for the BENNING CC 2 comprises:
3.1 One BENNING CC 2 with a rmly connected safety measuring line with
90° offset 4 mm safety connector,
3.2 One operating instructions manual
4. Description of appliance tester
The clamp-on adapter BENNING CC 2 is a measuring adapter for analog and
digital multimeters and is used for alternating current measurements up to
200 A.
See gure 1: Appliance front face
Moreover, the BENNING CC 2 current clamp adapter combined with the
BENNING IT 120 installation tester is appropriate for measuring load currents
from 0.5 A to 20 A.
The display and operator control elements specied in Fig. 1 are designated
as follows:
Measuring pliers, for clamping on the single wire alternating current
conductor.
Current probe bulge, protects against contact with conductor.
Opening lever, for opening and closing the measuring pliers.
Housing with label area
Safety measuring line with 4 mm safety connector, red, black, 90° offset.
5. General information
5.1 General details on the clamp-on adapter
Sensor type: Induction coil for measuring alternating current
5.1.1 Temperature coefcient of the measured value:
0.2 x (stated measuring precision)/ °C < 18 °C or > 28 °C, related to the
value for the reference temperature of 23° C.
5.1.2 Largest opening of pliers: 21 mm,
5.1.3 Maximum conductor diameter: 20 mm,
5.1.4 Appliance dimensions:
(L x W x H) = 135 x 50 x 30 mm
Appliance weight:
180 g
03/ 2007
BENNING CC 2
6
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning CC2 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning CC2 in the language / languages: English, German, Dutch, French, Italian, Polish, Swedish, Spanish as an attachment in your email.

The manual is 0,83 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info