English 4
Do not simultaneously touch the DC jack and the patient. That may result in
electrical shock.
To measure blood pressure, the arm must be squeezed by the cuff hard
enough to cause some numbness and possibly a temporary red mark to
the arm.
Follow local instructions specified in the hospital when the cuff is used on
several or infectious patients. Otherwise cross infection may result.
If the patient has a very weak or irregular heart beat, the device may have
difficulty in determining the blood pressure.
Should the battery short-circuit, it may become hot and potentially cause
burns.
Note
Do not modify the device.
The patient should be relaxed and avoid moving or talking during
measurement. Otherwise that may result in a measurement error.
To ensure accurate measuring, we recommend measuring the blood
pressure after being in a relaxed state for at least five minutes.
Care for after use
When the cuff is infected by blood or body fluid, it should be safely
disposed of according to local instructions or protocol to avoid any potential
spread of infectious disease.
Clean the device and cuff with a dry, soft cloth or a cloth dampened with
water and a neutral detergent. Never use benzene, thinner or other harsh
chemical to clean the device. For full details please read page 28.
When carrying out maintenance on the device, turn the power off and
remove the power cable from the outlet to prevent a risk of electrical shock.
Do not spray, do not pour or do not spill a liquid on the main body,
accessories, connectors, buttons or outlet ports.
Do not perform autoclave or gas sterilization (EOG, formaldehyde gas or
high concentration ozone, etc.) on the device as this could result in
degradation.
The user (Hospital, clinic, etc.) should have the management responsibility
for a use and maintenance for the medical electronic device. Be sure to
perform the specified daily and maintenance inspection for safe use.
Specified battery pack
Only the specified battery pack is allowed to be used with this device.
Used equipment, parts and battery are not treated as ordinary household
waste, and must be disposed of according to the applicable local
regulations.
Be sure to remove the specified AC adapter from the device when the
specified battery pack is being re-installed in the device. Otherwise that
may result in an electrical shock.
Remove the specified battery pack from the device, and keep it elsewhere
if you are not going to use the device for a month or more. Recharge the
battery once every six months. Otherwise the battery may degrade.
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AnD UM-211 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AnD UM-211 in the language / languages: English as an attachment in your email.

The manual is 0,62 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info