574293
4
Zoom out
Zoom in
Previous page
1/6
Next page
MPT-250B SPECIFICATIONS AND
OPERATING INSTRUCTIONS
DATASHEET
______________________________________________________________________
Daniels Manufacturing Corporation, 526 Thorpe Road, Orlando, FL 32824, U.S.A.
Phone: (407) 855-6161 Fax (407) 855-6884 www.dmctools.com Email: dmc@dmctools.com
Copyright©2013 All Rights Reserved Page 4 of 6 Rev. M 10/13 MPT-250B-DS
3.5 SIDE PANEL DIP SWITCH SETTINGS
The dip switches are accessed through the opening on the left side of the MPT-250B. The access opening is covered
with a plate that can be swung out of the way. Turn the large captivated screw counter clockwise until it comes loose
from the side cover. The label below the cover depicts the switch banks and lists the switch functions. The detailed
description of the switch settings is as follows:
NOTE: The unit must be turned off, and then back on for dip switch changes to take effect.
Bank SW1 (Toward rear of tester)
Switch 1: Not Used
Switch 2: Not Used
Switch 3: Not Used
Switch 4: Not Used
Switch 5: Factory Setting DO NOT adjust.
Switch 6: Down = Normal
Up = Set point measure enabled
Switch 7: Auto Stop Down=Off
Up=On
Bank SW2 (Toward front of tester)
Switch 1: Not Used
Switch 2: lb units Down
(Section 3.5.2) Kg units Up
N units Up
Switch 3: lb units Up
(Section 3.5.2) Kg units Down
N units Up
Switch 4: Factory Setting DO NOT adjust.
Switch 5: Factory Setting DO NOT adjust.
Switch 6: Factory Setting DO NOT adjust.
Switch 7: Factory Setting DO NOT adjust.
3.5.1 AUTO STOP
This feature stops the pull test when the piece under test fails. When the force applied to the piece drops below
70% of the displayed reading, the test is completed and the unit stops. This feature is turned on or off using
switch 7 of dip switch bank SW1.
ON=AUTO STOP enabled.
OFF=AUTO STOP disabled (the STOP switch must be used at the completion of a test).
NOTE: The AUTOSTOP function will not operate if the test range is at or below approximately 1.5lb. This is
necessary to allow for the preload that an operator will normally apply when inserting a connector for the
test. If the preload is greater than the lower limit, the unit will not start. In this case depress the reset
toggle switch next to the display to be able to begin test.
3.5.2 POUND, KILOGRAM, NEWTON
Dip switch 2 & 3 of dip switch bank SW2 control the type of units that are measured by the tester. If pounds are
desired switch 2 is down and switch 3 is up. If kilograms are desired switch 2 is up and switch 3 is down. If
Newton units are desired both switch 2 and 3 are up.
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Alphatron MPT-250B at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Alphatron MPT-250B in the language / languages: English as an attachment in your email.

The manual is 0,3 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info