Aerogen® Solo System Instruction Manual 23
10. To stop the nebuliser at any time, press the On/Off power button. The
indicator turns off to indicate that nebulisation has stopped.
Aerogen’s recommended input rate of medication into the Aerogen Solo
nebuliser during continuous nebulisation is up to a maximum of 12 mL per hour.
The upper limit of 12 mL per hour is based on Aerogen’s specification for the
minimum nebuliser flow rate. For directions on determining flow rates, refer
to the Optional Flow Rate Calculation method in the Functional Test section,
page25.
Warnings Specific to the Continuous Nebulization Tube Set
It is important to ensure that the maximum flow rate through the tube set
into the nebuliser must not exceed the output rate of the nebuliser.
Check for leaks from the system prior to and during use.
The graduations on the syringe are for indication use only.
Store at room temperature and use product within labelled shelf life.
To ensure correct and safe connection between the nebuliser and the
medication reservoir, trace the medication tube from the nebuliser back to
the medication reservoir to make sure the medication tube is connected to
the correct source.
The recommended syringe pump software setting with the Aerogen syringe
is typically the “60mL BD Plastipak” setting. This must be validated locally
before use. Refer to pump manual or manufacturer for guidance. These
pumps may also be used in accordance with local hospital or ward policies.
Ensure that the tethered silicone plug is attached to the Aerogen Solo when
connecting tube set.
Ensure that the tubing is safely orientated to prevent a trip hazard.
Rising level of medication in the reservoir may occur if the Aerogen Solo
nebuliser is turned off while the feed system is still on or the nebuliser is not
in its recommended orientation.
The level of the medication in the reservoir of the Aerogen Solo nebuliser
should be periodically monitored to ensure that the fill rate of medication
does not exceed the output rate of the nebuliser. A rising level of medication
in the reservoir indicates that the fill rate is exceeding the output rate of the
nebuliser.
Replace both the tube set and syringe when changing the type of medication.
If the syringe needs to be replaced during use (even when empty), turn off
the syringe pump and disconnect the nebuliser end of the tube set first.
Failure to do this may result in primed medication in the tube flowing into
the nebuliser reservoir.
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Aerogen Solo at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Aerogen Solo in the language / languages: English as an attachment in your email.

The manual is 5.4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Aerogen Solo

Aerogen Solo User Manual - German - 39 pages

Aerogen Solo User Manual - Dutch - 42 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info