Aerogen®
22
ASyringe Cap
Syringe
Tubing
(Syringe End)
Tubing
Tethered Silicone Plug
Tubing
(Nebuliser End)
Figure 17. Continuous Nebulization Tube Set
1. Ensure the Aerogen Solo nebuliser is firmly fitted into the Aerogen Solo
T-piece in the breathing circuit.
2. Remove the syringe cap from the medication-filled syringe.
3. Attach the syringe end of the tubing onto the syringe.
4. Prime the tubing until the medication reaches end of tubing (Point A).
Note: The tubing priming volume is maximum 3.65 mL.
5. Unplug the tethered silicone plug from the Aerogen Solo nebuliser, but do
not remove it from the nebuliser.
6. Screw the nebuliser end of the tubing onto the top of the nebuliser.
7. Insert the syringe filled with medication into the syringe infusion pump
(pump not shown in Figure 17) and set the appropriate flow rate (refer to
pump manual or manufacturer for guidance).
8. To start a continuous nebulisation cycle, press and hold the blue On/Off
power button from the off state for at least three seconds. Verify the green,
‘Continuous Nebulisation’ indicator light is on (Figure 4).
9. Observe nebuliser for correct operation. During continuous nebulisation,
the nebuliser is on continuously and the medication is nebulised on a drop
by drop basis. Nebulisation should be visible with regular intermittent
pauses. Medication level in the nebuliser reservoir should not rise during
use.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Aerogen Solo at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Aerogen Solo in the language / languages: English as an attachment in your email.

The manual is 5.4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Aerogen Solo

Aerogen Solo User Manual - German - 39 pages

Aerogen Solo User Manual - Dutch - 42 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info