EMI/EMC
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
The AV500 is intended for use in the electromagnetic environment specied below. The customer or the user of the AV500 should assure that it is used in such an environment.
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC
61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete, or ceramic tile. If oors are covered with synthetic
material, the relative humidity should be at least 30 %.
Electrical fast transient/burst IEC
61000-4-4
±2kV for power supply
lines
±1kV for input/output
lines
±2kV for power supply
Lines
±1kV for input/output
lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV line(s) to line(s)
±2kV line(s) to earth
±1kV line(s) to line(s)
±2kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
<5% UT >95% dip for
0.5 cycle
40% UT 60% dips
70% UT 30% dip
Mains power quality should be that of a typical commercial or hospital environment. If
the user of the AV500 requires continued operation during power mains interruptions, it
is recommended that the AV500 be powered from an uninterruptible power supply or a
battery.
Power frequency (50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be at levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The AV500 is intended for use in the electromagnetic environment specied below. The customer or the user of the AV500 should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms Portable and mobile RF communications equipment should be used no closer to any part of the AV500,
including cables, than the recommended separation distance calculated from the equation applicable to the
frequency of the transmitter. Recommended separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m). Field strengths from xed RF
transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each
frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects, and people.
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be
considered. If the measured eld strength in the location in which the AV500 is used exceeds the applicable RF compliance level above the AV500 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the AV500. Over the frequency range 150 kHz to 80 MHz,
eld strengths should be less than 3 V/m.
12
12


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AccuVein AV500 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AccuVein AV500 in the language / languages: English as an attachment in your email.

The manual is 1.23 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of AccuVein AV500

AccuVein AV500 User Manual - German - 16 pages

AccuVein AV500 User Manual - Dutch - 16 pages

AccuVein AV500 User Manual - French - 16 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info