Recommended separation distances between portable and mobile RF
communications equipment and the Emfit Tonic-Clonic Seizure Monitor.
The Emfit Tonic-Clonic Seizure Monitor is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the Emfit epileptic seizure alarm can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Emfit Tonic-Clonic Seizure Monitor alarm as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of
transmitter
150 kHz 80 MHz
d =1,2P
80 MHz 800 MHz
d =1,2P
800 MHz 2.5 GHz
d =1,2P
0,01
0,12 m
0,12 m
0,23 m
0,1
0,38 m
0,38 m
0,73 m
1
1,2 m
1,2 m
2,3 m
10
3,8 m
3,8 m
7,3 m
100
12 m
12 m
23 m
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
25. Manufacturer
Abilia AB
Råsundavägen 6
169 67 Solna, Sweden
Phone: +46 8 594 694 00
Email: info@abilia.se
Internet: www.abilia.com
31


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464031 Emfit Monitor at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464031 Emfit Monitor in the language / languages: English as an attachment in your email.

The manual is 1,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Abilia 464031 Emfit Monitor

Abilia 464031 Emfit Monitor User Manual - German - 36 pages

Abilia 464031 Emfit Monitor User Manual - Dutch - 36 pages

Abilia 464031 Emfit Monitor User Manual - French - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info