Guidance and manufacturer’s declaration – electromagnetic immunity
The Emfit Tonic-Clonic Seizure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the equipment should
assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
IEC-60601-
1-2 test level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines ±1 kV for
input/output
lines
IEC-60601-
1-2 test level
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV line(s)
to line(s)
IEC-60601-
1-2 test level
Mains power quality should be that of a
typical commercial or hospital
environment
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5 % UT (>95
% dip in UT)
for 0,5 cycle
40 % UT (60
% dip in UT)
for 5 cycles
70 % UT (30
% dip in UT)
for 25 cycles
<5 % UT (>95
% dip in UT)
for 5 sec
IEC-60601-
1-2 test level
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Emfit
epileptic seizure alarm enquires
continued operation during power
mains interruptions, it is recommended
that the Emfit Tonic-Clonic Seizure
Monitor be powered from an
uninterruptible power supply or battery
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
IEC-60601-
1-2 test level
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464031 Emfit Monitor at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464031 Emfit Monitor in the language / languages: English as an attachment in your email.

The manual is 1,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Abilia 464031 Emfit Monitor

Abilia 464031 Emfit Monitor User Manual - German - 36 pages

Abilia 464031 Emfit Monitor User Manual - Dutch - 36 pages

Abilia 464031 Emfit Monitor User Manual - French - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info