540488
72
Zoom out
Zoom in
Previous page
1/84
Next page
9-4
IM 1F5B2-01E
Table 9.2-2 Pressure Test Value
Pressure
2.1 MPa (21 kgf/cm
2
)
5.0 MPa (51 kgf/cm
2
)
2.9 MPa (29 kgf/cm
2
)
Flange Rating
JIS 10K
JIS 20K
ANSI Class 150
7.5 MPa (76 kgf/cm
2
)ANSI Class 300
9.2 Model and Suffix Codes
Model and Suffix Codes (Style Code: S2)
UYF200 Ultrasonic Vortex Flowmeter (Integral type, Remote type detector)
Model
UYF201J
UYF202J
UYF204J
UYF205J
UYF208J
UYF210J
UYF215J
UYF220J
.....................
.....................
.....................
.....................
.....................
.....................
.....................
.....................
15 mm (1/2inch)
25 mm (1 inch)
40 mm (1-1/2 inch)
50 mm (2 inch)
80 mm (3 inch)
100 mm (4 inch)
150 mm (6 inch)
200 mm (8 inch)
Converter –A
................
–N
................
Integral Type
Remote Type
Output Signal D
...............
N
...............
4 to 20 mA DC, Pulse,BRAIN
communication
Remote Type
Body Material –S
...........
Casting stainless steel (SCS 14A)
Process
Connection
(Note 1)
J1
........
J2
........
A1
.......
A2
.......
K1
.......
K2
.......
B1
........
B2
........
H1
........
H2
........
JIS10K Flange
JIS20K Flange
ANSI Class 150 Flange
ANSI Class 300 Flange
JIS10K Wafer
JIS20K Wafer
ANSI Class 150 Wafer
ANSI Class 300 Wafer
JPI Class 150 Wafer
JPI Class 300 Wafer
Electrical
Connection
0
........
2
........
3
........
4
........
JIS G1/2 Female
ANSI 1/2 NPT Female
DIN Pg 13.5 Female
ISO M20X1.5 Female
Indicator/
Totalizer (Note 2)
–D
...
–N
...
With Indicator/Totalizer
None
Options /
Refer to table Option Specifications
Suffix Code
Description
Electrical
Classification
/JF1
/FF1
/FS1
/KF1
/KS1
/CS1
JIS Explosionproof
FM Explosionproof
FM Intrinsic Safety
CENELEC Explosionproof
CENELEC Intrinsic Safety
CSA Intrinsic Safety
Table 9.2-3 Flowmeter Selection Guide
200
(8)
NO
YES
150
(6)
NO
YES
100
(4)
YES
YES
80
(3)
YES
YES
50
(2)
YES
YES
40
(1-1/2)
YES
YES
25
(1)
YES
YES
Process
Connection
Wafer
Type
Flange
Type
Body
Material
Casting
Stainless
Steel
Nominal Size mm (inch)
72


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Yokogawa UYFA21 Ultrasonic Vortex Flowmeter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Yokogawa UYFA21 Ultrasonic Vortex Flowmeter in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info