8-15
IM 1F5B2-01E
8. MAINTENANCE
8.4.3 Instructions on Use of Check Terminals of Amplifier Unit
Test all the check items hereinafter discussed in this subsection at the test points
between terminals P and COM and between terminals TP2 and COM.
CAUTION
Do not ground any test-purpose measuring instruments.
COM terminal
P termina
TP2 terminal
HHT terminal
Figure 8.4-1 Check Terminals of Amplifier Unit
(1) TP2 Terminal
Used for trigger level adjustment. See Subsection 8.2.3.
Using an oscilloscope, observe the voltage waveform appearing across terminals TP2
and COM. Make sure the ultrasonic emitter/detector of the sensor assembly and
related circuits are operating normally.
a.A waveform with the vortex frequency shown in Figure 8.4-2 should be observed at
this test point. The waveform should be neither at the power supply frequency nor
at a frequency at which the pipeline vibrates.
b.If piping conditions are inferior, the waveform should look like the one in Figure
8.4-3.
Figure 8.4-2 Normal Waveform at Terminal TP2
67


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Yokogawa UYF200 Ultrasonic Vortex Flowmeter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Yokogawa UYF200 Ultrasonic Vortex Flowmeter in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info