Item Specification
Report formats, external
printer*
Auto ECGs
Average cycles
3x4 @ 25 mm/s, 3x4 @ 50 mm/s, 3x4 + 1R @ 25 mm/s, 3x4 + 3R @ 25 mm/s,
3x4 - 5.0s @ 25 mm/s, 3x4 - 5.0s @ 50 mm/s, 6x2 - 5.0s @ 25 mm/s,
12x1 - 5.0s @ 25 mm/s, No print
3x4 + 3R @ 25 mm/s, 3x4 + 3R @ 50 mm/s, 2x6 + 1R @ 25 mm/s,
2x6 + 1R @ 50 mm/s, 3x4 + 2R @ 25 mm/s, 3x4 + 2R @ 50 mm/s,
6x2 + 2R @ 50 mm/s, No print
ECG storage (in test
directory)
At least 25 ECG tests
Patient storage (in
worklist, CP 50 Plus only)
Up to 50 patients
Frequency range 0.3 to 150 Hz
Digital sampling rate > 1,000 samples/second/channel
Pacemaker detection ANSI/AAMI EC11
Power requirement Universal AC power supply 110-240 V~, 50/60 Hz ~, 1.5 A maximum
AC fuses Time-lag type, 2.0-amp 250-V rating, Littlefuse 0218002P or equivalent
Rechargeable battery 10.8V, 1.9Ah minimum, 3-cells Li-ion “smart” battery. A fully charged battery prints up to 25
ECGs with one extra copy per test.
Filters
High-performance
baseline
Muscle tremor
AC interference
0.5 Hz
35 Hz
50 Hz or 60 Hz
Safety, EMC, and
regulatory compliance
ANSI/AAMI EC11**
CAN/CSA C22.2 No. 601.1
CAN/CSA C22.2 No. 601.1.1
CAN/CSA C22.2 No. 601.1.2
CAN/CSA C22.2 No. 601.1.4
CAN/CSA C22.2 No. 601.2.25
UL60601-1
IEC/EN 60601-1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-1-6
IEC/EN 60601-2-25 ***
IEC/EN 60601-2-51 (2x6 lead arrangement)
Standard connectivity Mini USB client and USB host, Ethernet (optional - available with CP 50 Plus only)
Connectivity with
electronic medical records
(CP 50 Plus only)
Through the Welch Allyn CardioPerfect workstation software
44 Specifications CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph
48


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn CP 50 Plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn CP 50 Plus in the language / languages: English as an attachment in your email.

The manual is 1,34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn CP 50 Plus

Welch Allyn CP 50 Plus User Manual - Dutch - 52 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info