Electromagnetic emissions
Harmonic emissions
IEC 61000-3-2
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Electromagnetic immunity
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn CareTemp™ Touch Free Thermometer should assure that it is used
in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in
U
T
) for 0.5 cycle
40 % U
T
(60 % dip in
U
T
) for 5 cycles
70 % U
T
(30 % dip in
U
T
) for 25 cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 s
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Electromagnetic immunity
The Welch Allyn CareTemp™ Touch Free Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn CareTemp™ Touch Free Thermometer should assure that it is used
in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
20
Guidance and manufacturer's declaration CareTemp™ Touch Free Thermometer
24


Need help? Post your question in this forum.

Forumrules
1

Forum

welch-allyn-caretemp

Reset search

  • the thermometer will not take a temperature, it shows ERR on the screen, it is only 8 months old. Submitted on 15-4-2021 at 14:55

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn CareTemp at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn CareTemp in the language / languages: English as an attachment in your email.

The manual is 0,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn CareTemp

Welch Allyn CareTemp User Manual - German - 30 pages

Welch Allyn CareTemp User Manual - Dutch - 30 pages

Welch Allyn CareTemp User Manual - French - 30 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info