536170
23
Zoom out
Zoom in
Previous page
1/36
Next page
50
Mode de service Gamme de mesure Précision Résolution
Tension 400 mV ±(1,0% + 5 ch) 0,1 mV
continue 4 V ±(0,9% + 5 ch) 1 mV
40 V ±(0,9% + 5 ch) 10 mV
400 V ±(0,9% + 5 ch) 100 mV
1000 V ±(1,4% + 5 ch) 1 V
Protection de surcharge : 1000 VDC pendant moins de 10 s
Tension Non spécifiée pour des tensions d’entrée < 50mV
alternative 4 V ±(1,5%+5 ch) 1 mV
40 V ±(1,5%+5 ch) 10 mV
400 V ±(1,0%+5 ch) 100 mV
750 V ±(2%+5 ch) 1 V
Résistance d’entrée > 10 Mohms, fréquence de la tension alternative : 50 Hz à 400 Hz
Protection de surcharge : 750 VACrms < à 10 s
Affichage sur l’appareil VC-840 comme valeur efficace vraie (True Rms), facteur
Crest maxi. 3
Courant continu 400 µA ±(1,0%+5 ch) 0,1 µA
4000 µA ±(1,0%+5 ch) 1µA
40 mA ±(1,2%+5 ch) 0,01mA
400 mA ±(1,2%+5 ch) 0,1mA
4 A ±(1,5%+5 ch) 0,001A
20 A ±(1,5%+5 ch) 0,01A
Protection de surcharge, voir ci-après
Courant alternatif 400 µA ±(2,0%+5 ch) 0,1µA
4000 µA ±(2,0%+5 ch) 1µA
40 mA ±(2,0%+5 ch) 0,01mA
400 mA ±(2,0%+5 ch) 0,1mA
4 A ±(2,5%+5 ch) 0,001A
20 A ±(2,5%+5 ch) 0,01A
Fréquence du courant alternatif : 50Hz à 400 Hz
Protection de surcharge, voir ci-après
Affichage sur l’appareil VC-840 comme valeur efficace vraie (True Rms), facteur
Crest maxi. 3
Résistance 400 ohms ±(1,2%+2 ch) 0,1 ohm
4 k ohms ±(1,0%+2 ch) 0,001kohm
40 k ohms ±(1,0%+2 ch) 0,01kohm
400 k ohms ±(1,0%+2 ch) 0,1kohm
4 Mohms ±(1,2%+2 ch) 0,001Mohm
40 Mohms ±(1,5%+2 ch) 0,01Mohm
Contrôleur de continuité: Signal acoustique pour les résistances
< env. 70 ohms
Capacité 40 nF ±(3,0%+10 ch) 0,01nF
C 400 nF ±(3,0%+5 ch) 0,1nF
4 µF ±(3,0%+5 ch) 0,001µF
40 µF ±(3,0%+5 ch) 0,01µF
100 µF ±(4,0%+5 ch) 0,1µF
23
- Always make sure before measuring voltages that another measuring function is
not active (resistance measurement, diode test etc.).
- Before changing the measuring range, the test prods or adapters have to be
removed from the object to be measured.
- Check your measurement instrument or your measuring lines and adapters for
their correct functioning or for damage(s) before starting a measurement.
- Do not use the measuring instrument in rooms or at unfavourable ambient condi-
tions in which there may or could be combustible gases, vapours or dust. For your
own safety, always ensure that the measuring instrument and cables cannot
become wet or damp. Avoid operation in the immediate vicinity of:
a) magnetic and electromagnetic fields (transformers, motors, coils, relays, contac-
tors, electromagnets etc.)
b)electrostatic fields (charges/discharges)
These could falsify the value measured.
- For safety reasons, when measuring only use the supplied measuring cables
which are approved for the specifications of the measuring instrument. These are
the only cables permitted for use.
- To avoid an electric shock make sure that you do not touch the test prods or the
connections to be measured (measuring points), even indirectly, during measure-
ment.
- The voltage between any socket of the measuring instrument and the ground may
not be higher than 1000V CAT III / 600V CAT IV.
Handling and Starting-up Operation
A Display indications and symbols for the operating modes
„HOLD“ stands for Data Hold; the measuring value is kept (e.g. for the record)
until the HOLD button is pressed anew.
„REL“ stands for relative value measurement (= reference measurement)
„RANGE“ stands for (measuring) range; for the manual range selection.
„AUTO“ is indicated in the display (black background), if the automatic range
selection is active.
„O.L.“ stands for overload. The measuring value is too high or the measuring
range is exceeded
Battery replacement symbol
Symbol for the diode test
Symbol for the acoustic continuity check
all other symbols standing for the different measuring units:
AC = alternating quantity
DC = direct quantity
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Voltcraft VC960 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Voltcraft VC960 in the language / languages: English, German, Dutch, French as an attachment in your email.

The manual is 0,61 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info