To turn off this function, press the MAX/MIN (View/) button for approx. 2 s.
The MAX/MIN function is not available in all measuring functions.
16. LOW IMP. 400 K FUNCTION
This function may only be used for voltages of a max. of 1000 V and a max. of 3
seconds!
This function allows the reduction of the measuring impedance from 10 M to 400 k in the voltage
measuring range. Reducing the measuring impedance suppresses possible phantom voltages which
could falsify the measuring result.
Press the
-button (9) during the voltage measurement (max. 1000 V!) for a max. 3 seconds. After
release, the multimeter has the normal measuring impedance of 10 M. While the button is pressed,
a signal sounds and
appears on the display.
17. COMPARISON MODE (COMP MODE)
In the comparative measurement, once an upper and lower limit of the current measured parameter
is set, the main display will show only the actual status of the current measured value compared to
the preset range limit.
Proceed as follows to set the limit values for comparison mode:
1. Turn on the DMM and select the appropriate measurement range.
2. Insert the test leads into the correct measuring sockets.
3. Hold down the SHIFT/SETUP button (3) for approx. 2 s.
will appear on the display. You
are now in SHIFT mode.
4. Hold down the SHIFT/SETUP button again for approx. 2 s to enter the SETUP menu.
5. To set the upper limit for the comparison mode, press the SHIFT/SETUP button until the menu
item “COMP MAX” is selected. An asterisk symbol * to the left of the menu item shows that the
menu item is selected.
6. You can now select the individual number of digits using the REL/PC (Log/) and MAX/MIN
(View/) buttons and change their values with the H/LIGHT (+/COMP) or RANGE (-) button.
105
h) Mesure de capacité
Assurez-vous que tous les éléments de circuit, les circuits et les composants à
mesurer, ainsi que les autres objets de mesure sont bien hors tension.
Respectez scrupuleusement la polarité des condensateurs électrolytiques.
1. Allumez le DDM et sélectionnez la plage de mesure «
».
2. Insérez le cordon de mesure rouge dans la douille de mesure
°CHzV (7), le cordon de mesure noir dans la douille de
mesure COM (8) (Fig. 8).
3. L’écran af che l’unité « nF ».
En raison de l’entrée de mesure sensible, une valeur peut
s’af cher à l’écran lors de cordons de mesure « ouverts ».
En appuyant sur la touche REL/PC (Log/) (10c), l’écran
est réglé sur « 0 ». La valeur de base (différence relative)
est af chée à côté du symbole delta
. La fonction
Autorange est désactivée.
4. Reliez à présent les deux pointes de sonde (rouge = pôle
positif / noir = pôle négatif) à l’objet à mesurer (condensateur).
À l’écran, la capacité s’af che rapidement. Attendez que
la valeur af chée se soit stabilisée. En cas de capacités
>400 F, cela peut durer quelques secondes.
Fig. 8
5. L´af chage « OL » (pour Overload = dépassement) à l´écran indique que vous avez dépassé la
plage de mesure.
6. À la n de la mesure, retirez les cordons de mesure de l’objet à mesurer et éteignez le DMM.
66


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Voltcraft VC-890 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Voltcraft VC-890 in the language / languages: English, German, Dutch, French as an attachment in your email.

The manual is 6,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info