Use the “Hz%” button (11) to switch the display to frequency (Hz) and pulse length (%).
Each press switches the function. Autorange is deactivated. This enables monitoring of
possible ripple voltage (superimposed alternate voltage). In case of pure direct voltage, the
frequency and % displays are “zero”. Without a measured signal, phantom values may be
displayed. They disappear when a direct voltage is measured.
c) Current Measuring “A“ 10
ƽ
The voltage in the measuring circuit may not exceed 600 V.
Do not measure any currents above 10 A in the 10Arange and no currents above 400
mAin the µA/mA range, otherwise the fuses trigger.
Measuring >5 A may only be performed for max. 30 seconds and at 15 minute inter-
vals.
Always start current measurements at the highest measurement range and switch down to lower ranges
if necessary. Before changing the measurement range, always shut off the circuit. All current measuring
ranges are secured with fuses and thus protected against overload.
Proceed as follows to measure “A-DC” direct current:
- Turn the DMM on and select measuring function “A”.
The table shows the different measuring functions and
possible measuring ranges. Select your measuring range
and the respective measuring jacks. The display indicates
“DC”.
Measuring Measuring Measuring
function range jacks
µA <4000 µA COM + mAµA
mA 4mA– 399 mA COM + mAµA
10A 400 mA– 10 A COM + 10A
- Insert the red measuring line into the µA/mAor 10Amea-
suring jack. Plug the black measuring line into the COM
socket.
- Now connect the two measuring prods in series to the object to be measured (battery, switch etc.).
The polarity of the respective measured value is indicated on the display together with the measured
value.
When a minus “-” appears in front of the measured value when measuring DC, the
measured voltage is negative (or the measuring lines have been swapped).
After measuring, remove the measuring lines from the measured object and turn the DMM off.
44


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Voltcraft MT-52 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Voltcraft MT-52 in the language / languages: English, German, Dutch, French as an attachment in your email.

The manual is 2,58 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info