Performing the RGB calibration.
To ensure the accuracy of each measuring operation, the measuring device should be regularly calibrat-
ed using the enclosed calibration card.
Calibration is only possible in RGB mode and only with non-illuminating surfaces.
The sensor illumination must be activated.
Proceed as follows for calibration:
Connect the sensor to the measuring device and turn it on.
Press the measuring button (4 or 16) and check the activated sensor illumination. If this is not lit, turn
the illumination in the sensor head on via the button „LIGHT ON/OFF“ (9).
Open the envelope with the white calibration card and place the sensor head (15) in the center of the
white calibration circle. The sensor head must be plane.
Press the two buttons „Function“ (7) and „CAL“ (2) simultaneously. The calibration procedure starts and
the display shows the following:
The start screen appears on the display for approx. 30 seconds. The sen-
sor may not be moved or lifted during the entire calibration procedure.
After this time, the first calibration value appears on the display.
The second calibration value appears after approx. 10 seconds.
Finally, the third calibration value is displayed after another 10 seconds.
The end of the calibration procedure is signalled by three beeps. The CAL
symbol disappears after this procedure. Now the measuring device is
ready for further measuring operations.
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Voltcraft 122953 - Colour Analysis at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Voltcraft 122953 - Colour Analysis in the language / languages: English, German, Dutch, French as an attachment in your email.

The manual is 0,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info