514911
33
Zoom out
Zoom in
Previous page
1/40
Next page
Express
2012-03
Page 31
Higher performance and/or high utilization (daily use or use on difficult terrain), increase the
inspection intervals for wear parts (carbon brushes, pollution motor, tyre profile, ...) to at least
half-yearly review.
Checking measurements may only be carried out by skilled persons trained on the
wheelchair and at least under the supervision of an electrician who knows the checking
instruments and processes. Only an electrician is able to release the electric wheelchair for
use after performing measurements or servicing.
The service must only be confirmed in the maintenance plan if a minimum of all the above-
mentioned aspects have been checked.
4.5 Troubleshooting
The following list will assist you with the troubleshooting of your electric wheelchair.
Problem Cause of problem
The electric wheelchair refuses to go
once started.
No indication of charge level
Electric wheelchair is not switched on.
Battery plug not connected (batteries not making contact).
Thermal safety fuses disengaged. (check that the fuses of the
operator unit and the batteries are defective or have burn out)
Batteries defective (totally discharged).
Defective operator unit.
Loom of cables defective.
The electric wheelchair refuses to go
once started. Charge level insufficient.
Motor/gears in neutral.
Magnetic brake defective.
Motor defective.
Operating error
Possible safety shut down or operating error (joystick).
Put the joystick in neutral position and restart the system.
Thermal safety fuses disengage.
Motor overloaded (see “Technical specifications”).
Thermal safety fuses defective.
Battery error
Check batteries and cabling, if necessary charge.
If necessary, replace batteries.
Batteries cannot be charged.
Batteries not fitted correctly.
Loading socket defective
Wrong charger.
Loading device defective
4.6 Disinfection
L
LL
L
WARNING:
Dangerous Products - The use of disinfectants is restricted to
authorized personnel.
L
LL
L
WARNING:
Dangerous Products, risk of skin irritation - Wear suitable protective
clothing. Take note of the product information of the solutions concerned.
All parts of the wheelchair can be treated by scrubbing with a disinfectant.
All steps taken to disinfect rehabilitation equipment, their components or other accessory
parts are to be recorded in a disinfection report containing a minimum of the following
information (with product documentation appended):
Date of the
disinfection
Reason Specification Substance and
concentration
Signature
Table 2: Example of a disinfection book
Abbreviations used in column 2 (reason):
V = Suspected infection IF = Infection case W = Repetition I = Inspection
For a blank sheet of a disinfection book see § 9.
The recommended disinfectants for scrubbing (based on the list provided by the Robert Koch
Institute, RKI) are standing in the table below. The current state of the disinfectants included
in the RKI list can be obtained from the Robert Koch Institute (RKI) (homepage: www.rki.de).
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Vermeiren Express at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Vermeiren Express in the language / languages: English as an attachment in your email.

The manual is 1,49 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Vermeiren Express

Vermeiren Express User Manual - Dutch - 44 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info