4. Overvoltage/installation category
DMMs are categorized depending on the risk and severity of transient overvoltage
that might occur at the point of test. Transients are short-lived bursts of energy
induced in a system, e.g. caused by lightning strike on a power line.
The existing categories according EN 61010-1 are:
CAT I
A CAT I-rated meter is suitable for measurements on protected electronic
circuits that are not directly connected to mains power, e.g. electronics
circuits, control signals…
CAT II
A CAT II-rated meter is suitable for measurements in CAT I-environments
and mono-phase appliances that are connected to the mains by means of
a plug and circuits in a normal domestic environment, provided that the
circuit is at least 10m apart from a CAT III- or 20m apart from a CAT IV-
environment. E.g. household appliances, portable tools…
CAT III
A CAT III-rated meter is suitable for measurements in CAT I- and CAT II-
environments, as well as for measurements on (fixed) mono- or poly-
phased appliances which are at least 10m apart from of a CAT IV-
environment, and for measurements in or on distribution level equipment
(low-voltage distribution boards, lighting circuits, electric ovens).
CAT IV
A CAT IV-rated meter is suitable for measuring in CAT I-, CAT II- and CAT
III-environments as well as on the primary supply level.
Note that for all measurements on equipment for which the supply cables
run outdoors (either overhead or underground) a CAT IV meter must be
used.
Warning:
This device was designed in accordance with EN 61010-1 installation category CAT
IV 600V / CAT III 1000V. This implies that certain restrictions in use apply that are
related to voltages and voltage peaks which can occur within the environment of
use. Refer to the table above.
This device is suitable for measurements up to 1000V:
Protected electronic circuits that are not directly connected to mains power, e.g.
electronics circuits, control signals, circuits behind isolating transformer…
circuits that are directly connected to mains power, but limited to:
o measurements on mono-phase appliances that are connected to the mains by
means of a plug
o mono-phase appliances and circuits directly connected to the mains in a normal
domestic environment, provided that the circuit is at least 10m apart from a
CAT IV-environment. E.g. household appliances, portable tools, light circuits at
more than 10m from a distribution board
measurements in/on low-voltage distribution boards (distribution boards behind
meter box)
measurements on (fixed) mono- or poly-phased appliances and circuits except in
CAT IV-environments (e.g. mains outlets, electric ovens, lighting circuits, bus
bars, low-voltage distribution boards and circuit breakers).
This device is suitable for measurements up to 600V:
Measurements on distribution equipment and outdoor installations including meter
boxes and equipment/circuits outside or remote from the domestic environment
e.g. circuits in sheds, garden houses and free-standing garages , or circuits using
underground wiring e.g. garden lighting, pool-pump...
This device is NOT suitable for:
Voltages above 1000V
Currents above 10A
This device is only suitable for measurements up to 600V in
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Velleman DVM4100 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Velleman DVM4100 in the language / languages: English, German, Dutch, French, Spanish as an attachment in your email.

The manual is 1,07 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info