756403
4
Zoom out
Zoom in
Previous page
1/10
Next page
4 EN
pyrometer BP21
Information about the device
Device description
The pyrometer BP21 uses an infrared sensor to measure
surface temperatures without contact. An integrated dual laser
pointer can be switched on for the precise determination of the
measuring spot diameter.
In addition, you can set the degree of emission of the material to
be measured which will result in a more precise measurement
result.
The device further features an alarm function. When exceeding
or falling below the specified values, the device emits an
acoustic signal.
Moreover, the device can indicate either the highest or lowest
value of the measurement.
The display can be illuminated as needed. An automatic switch-
off function saves the battery when not in use.
Measuring principle
The device measures the temperature by means of an infrared
sensor. Important factors playing a role in the temperature
measurement are the measuring spot diameter and the degree
of emission.
Measuring spot
Observe the ratio of the distance(D) to the measuring spot
diameter(S). The larger the distance to the object, the larger the
measuring spot diameter and hence, the less precise the
measured result, for the device calculates an average
temperature from all the temperatures in the measuring spot.
Spot 12.5mm 100 mm
Distance 150 mm 1200 mm
D:S = 12:1 Laser
Infrared
Degree of emission
The degree of emission is used to describe the energy radiation
characteristics of a material.
Most organic materials have an emissivity of0.95. Metals or
shiny materials come with a much lower value.
A material's emissivity depends on various factors, e.g. on
Material composition
Surface condition
Temperature
The emissivity can range between 0.1 and 1 (in theory).
The following rule of thumb can be assumed:
When a material is rather dark and its surface texture matt,
it probably has a high emissivity.
The brighter and smoother the surface of a material, the
lower will be its emissivity, presumably.
The higher the degree of emission of the surface to be
measured, the better it is suited for non-contact
temperature measurement by use of a pyrometer or
thermal imaging camera, since falsifying temperature
reflections become negligible.
Nevertheless, entering a degree of emission as appropriate as
possible is indispensable for a precise measurement.
Material Degree of emission
Aluminium, roughened 0.1 to 0.3
Aluminium, alloy A3003,
oxidized
0.3
Aluminium, oxidized 0.2 to 0.4
Asbestos 0.92 to 0.95
Tarmac 0.92 to 0.95
Basalt 0.7
Concrete 0.92 to 0.95
Bitumen 0.98 to 1.00
Lead, oxidized 0.2 to 0.6
Lead, rough 0.4
Roofing felt 0.95
Ice 0.98
Iron (forged), blunt 0.9
Iron, oxidized 0.5 to 0.9
Iron, rusted 0.5 to 0.7
Enamel varnish, black 0.95
Earth 0.92 to 0.96
Paint (not alkaline) 0.90 to 0.95
Paint (non-metal) 0.95
Gypsum 0.60 to 0.95
Glass, pane 0.85 to 0.95
Rubber 0.92 to 0.95
4


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Trotec BP21 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Trotec BP21 in the language / languages: English as an attachment in your email.

The manual is 0,45 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Trotec BP21

Trotec BP21 User Manual - German - 10 pages

Trotec BP21 User Manual - Dutch - 10 pages

Trotec BP21 User Manual - French - 10 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info