BPM ARM 5500 XLD
USER GUIDE / HANDLEIDING / MANUEL D’UTILISATEUR
GEBRAUCHSANWEISUNG / MANUAL DE USUARIO
BRUKSANVISNING / BRUGERVEJLEDNING
ANVÄNDARHANDBOK / KÄYTTÖOHJE / MANUALE D’USO
MANUAL DO UTILIZADOR / UŽIVATELSKÁ PØÍRUÈKA
 /HASZNLATI TMUTAT
INSTRUKCJA OBSUGI/MANUAL DE INSTALARE SI UTILIZARE
UŽÍVATEL’SKÝ MANUÁL
V1.0 - 10/09
0197
The features described in this manual are published with reservation to modifications.
This Blood Pressure Monitor complies with the EC Directives and
bears the CE Mark "CE 0197".
This product is in compliance with the essential requirements and
other relevant provisions of the directive 93/42/EEC. The declaration
of conformity can be found on:
http://www.topcom.net/support/cedeclarations.php
The quality of the device has been verified and is in line with the
provisions of the EC council directive 93/42/EEC (EN IEC60601-1-
General requirements for safety, EN IEC 60601-1-2:2001 -
Electromagnetic compatibility-Requirements and tests) dated 14 June
1993 concerning medical devices and the EN performance standards
as follow:
EN 1060-1 Non-invasive sphygmomanometers - General
requirements
EN 1060-3 Non-invasive sphygmomanometers -
Supplementary requirements for electromechanical blood
pressure measuring systems.
EN 1060-4 Non-invasive sphygmomanometers - Test
procedures to determine the overall system accuracy of
automated non-invasive sphygmomanometers.
EC
REP
Authorized Representative in the European Community
Topcom Europe NV,
Grauwmeer 17
3001 Heverlee, Belgium,
Manufacturer
Andon Health Co., Ltd
7/F,Hua Qiao Chuang Ye Plaza,No.10 Jin Ping Rd,
Ya An St,Nankai Dist,Tianjin,China
3
ENGLISH
TOPCOM BPM Arm 5500 XLD
1 Introduction
Congratulations with your purchase of the Topcom BPM Arm 5500 XLD. This fully
automatic, Arm mounted blood pressure monitor is both easy to use and ideally suited
for daily measurements. The large display includes systolic, diastolic and pulse, all
clearly shown at the completion of each reading.
In addition, you can store up to 30 measurements per Memory Zone, ideal for users who
want to monitor and track their blood pressure on a regular basis. The BPM Arm 5500
XLD is compact and portable, making it ideal for home use and travel.
2 Safety instructions
This product is for household use only. It is not intended as a substitute for
consultation with your doctor.
For users diagnosed with common arrhythmia (atrial or ventricular premature beats
or atrial fibrillation), diabetes, poor circulation of blood, kidney problems, or for
users suffered from stroke, or for unconscious users, the device might not be
suitable for use. Consult your doctor in case of any doubt.
The unit should not be operated by children to avoid hazardous situations.
The unit contains high-precision assemblies. Therefore, avoid extreme
temperatures, humidity and direct sunlight. Avoid dropping or strongly shocking the
main unit, and protect it from dust.
Leaking batteries can damage the unit. Remove the batteries when the unit is not
used for a long time.
Do not press the start/stop button when the cuff is not placed around the
arm.
Do not disassemble the main unit or cuff.
If the unit is stored in a cold place, allow it to acclimatize at room temperature
before use.
Clean the blood pressure monitor body and the cuff carefully with a slightly damp,
soft cloth. Do not press. Do not bend the preformed cuff inside out. Do not wash the
cuff or use chemical cleaner on it. Never use thinner, alcohol or petrol (gasoline) as
cleaner.
Please read this manual carefully before use. For specific
information on your own blood pressure, contact your
physician. Put the user guide away in a safe place for
future reference.
9
To avoid inaccurate results caused by electromagnetic
interference between electrical and electronic equipments,
do not use the device near a cell phone or microwave oven.
4
TOPCOM BPM Arm 5500 XLD
3 Blood pressure
3.1 What is blood pressure?
Blood pressure is the pressure exerted on the artery tube while blood flows through the
arteries. The pressure measured when the heart contracts and sends blood out of the
heart is systolic (highest). The pressure measured when the heart dilates with blood
flowing back into the heart is called diastolic (lowest) blood pressure.
3.2 Why measure your blood pressure?
Among the various health problems afflicting modern people, problems associated with
high blood pressure are by far the most common. High blood pressure's dangerously
strong correlation with cardiovascular diseases and high morbidity has made measuring
blood pressures a necessity of identifying those at risk.
3.3 Blood pressure standard
The World Health Organization (WHO) and
National High Blood Pressure Education
Program has developed a Blood pressure
standard, according to which areas of low
and high-risk blood pressure are identified.
This standard however, is a general
guideline as individual's blood pressure
varies among different people and different
groups...etc.
The WHO classification indicator on your
BPM Arm 5500 XLD will indicate the blood
pressure class together with the result.
It is important that you consult your physician regularly. Your physician will tell you your
normal blood pressure range as well as the point at which you will be considered at risk.
Blood pressure
classification
Systolic
(mmHg)
Diastolic
(mmHg)
Color
indication
Stage 3 Hypertension >180 >110 Red
Stage 2 Hypertension 160-179 100-109 Orange
Stage 1 Hypertension 140-159 90-99 Yellow
High - Normal 130-139 85-89 Green
Normal 120-129 80-84 Green
Optimal <120 <80 Green
12
6
5
4
3
2
1
5
ENGLISH
TOPCOM BPM Arm 5500 XLD
3.4 Blood pressure fluctuation
Blood pressure fluctuates all the time!
You should not be worried if you encountered two or three measurements at high levels.
Blood pressure changes over the month and even throughout the day depending on
circumstances (state of mind, temperature, ...)
4 Battery installation
Open the battery compartment by pulling the
cover away on the bottom of the unit.
Insert 4 AA non - rechargeable batteries. Follow
the polarity as indicated in the battery
compartment.
5 Power adapter
The BPM Arm 5500 XLD can operate on
batteries or on the enclosed adapter (DC 6V).
Connect the power adapter jack with the
power adapter connector on the right
side of the unit .
Plug the power adapter into the wall
socket.
Do not short-circuit & dispose of in fire.
Remove the battery if this device is not
going to be used for a long period.
When the batteries are low, is
displayed. Low batteries must be
replaced.
7
4
4
Only use the enclosed power adapter
6
TOPCOM BPM Arm 5500 XLD
6 Buttons
1. Time
2. Date
3. Systolic pressure
4. Power adapter connector
5. Diastolic pressure
6. Heart rate
7. Battery low icon
8. Memory button
9. Start/stop button
10. Memory location
11. Irregular heartbeat icon
12. WHO classification indication
13. Arm cuff connector
DIA
mmHg
SYS
mmHg
PUL/min
M
1 2
10
11
12
89
7
3
5
6
4
13
7
ENGLISH
TOPCOM BPM Arm 5500 XLD
7 Applying the cuff
1. Remove all watches, jewelry, etc. prior to
attaching the cuff. Clothing sleeves should
be rolled up.
2. Press your brachial artery by two fingers
approximately 2,5 cm above the elbow on
the inside of your left arm to determine
where your strongest pulse is.
3. Slide the end of arm cuff furthest from the
tube through the metal ring to a loop. The
smooth cloth should be on the inside of the
cuff. The Velcro will be on the outside of the
cuff.
4. Put the left arm through the cuff loop. The
bottom of the cuff should be approx. 1,5 cm
above the elbow. The cuff tube should lie
over the brachial artery on the inside of the
arm.
5. Pull the cuff so that the top and bottom
edges are tightened around your arm.
6. When the cuff is positioned properly, press
the Velcro firmly against the pile side of the
cuff. The metal ring will not touch your skin.
7. Relax your entire body, especially the area
between your elbow and fingers. Place your
elbow on a table so that the cuff is at the
same level as your heart. Avoid leaning back
while the measurements being taken.
FP
8
TOPCOM BPM Arm 5500 XLD
8 Measurement
8.1 Important tips
This monitor automatically switches off 1 minute after the last key operation.
To interrupt the measurement, simply press any key. The cuff will deflate
immediately after a key is pressed.
Do not measure your blood pressure immediately after consuming a large meal. To
obtain more accurate readings, please wait one hour before measuring.
Do not smoke or drink alcohol before measuring your blood pressure.
You should not be physically tired or exhausted while taking a measurement.
It is important that you relax during measurement. Try to take 15-minutes rest
before a reading.
Do not take measurements if you are under stress or under tension.
Take your blood pressure at normal body temperature. If you are feeling cold or
hot, wait a while before taking a measurement.
If the monitor is stored at very low temperature (near freezing), have it placed at a
warm location for at least one hour before using it.
Wait about 5 minutes before taking the next pressure measurement.
During the measurement, do not talk or move your arm or hand muscles.
8.2 Procedure
Press the start/stop button to activate the device.
All segments will be displayed.
Once the actual Memory zone is displayed (U1-3), press
the memory button to select the desired memory
zone.
If no button is pressed within 5 seconds, the measurement
will start automatically.
The display will show the last measurement result before
inflating the cuff to the level that is right for you.
When the right level is reached, the cuff will deflate.
This monitor will re-inflate automatically to
approximately 220mmHg if the system detects that
your body needs more pressure to measure your
blood pressure.
When the measurement is completed, systolic, diastolic
and pulse will be shown simultaneously on the LCD
screen.
The blood pressure classification indicator will be
displayed.
When an irregular heartbeat is detected, ‘ is
displayed.
When a mistake was detected during measurement an
error ‘Er’ message is displayed (see “10 Troubleshoot” ).
Press the start/stop button to turn the unit off.
DIA
mmHg
SYS
mmHg
PUL/min
DIA
mmHg
SYS
mmHg
PUL/min
DIA
mmHg
SYS
mmHg
PUL/min
9
M
8
11
9
9
ENGLISH
TOPCOM BPM Arm 5500 XLD
9Memory
The BPM Arm 5500 XLD has 3 Memory zones. Each can store
up to 30 measurements including the date and time.
The result will be stored automatically in the selected Memory
Zone after the measurement.
Select the desired Memory Zone before measurement after
pressing the start/stop button .
To recall measurement results:
When the clock is displayed,
Press the memory button to enter memory mode.
Press the start/stop button select the desired
memory zone. The number of stored measurement results
is display.
Press the memory button . The average of all
results in the selected memory zone will be displayed for 3
seconds before displaying the last measurement result.
Press the memory button repeatedly for prior
values stored.
After the oldest record is displayed, the unit will switch
back to normal time.
To erase all data from one memory zone:
Press the memory button to enter memory mode.
Press the start/stop button to select the desired
memory zone.
When the last measurements result is displayed, press
and hold the memory button for 3 seconds to
erase all records from the selected memory zone.
DIA
mmHg
SYS
mmHg
PUL/min
DIA
mmHg
SYS
mmHg
PUL/min
DIA
mmHg
SYS
mmHg
PUL/min
9
M
8
9
M
8
M
8
M
8
9
M
8
10
TOPCOM BPM Arm 5500 XLD
9.1 Time adjustment
10 Troubleshoot
To adjust the date an time:
Press the start/stop and memory button
simultaneously. The month will blink on the display.
Press the memory button successively to change
the month.
Press the start/stop button to confirm the month.
The day will blink on the display.
Change the day, hours and minutes as described above,
using the memory button to change and the start/
stop button to confirm every setting.
After setting the minutes, the device is ready to use.
Error message
on display
Reason Solution
Batteries low Change the batteries
Er0 Pressure system is unstable before
measurement.
Don't move and try again.
Er1 Systolic pressure detect fail
Er2 Diastolic pressure detect fail
Er3 Pneumatic system blocked or cuff is
too tight during inflation
Place the cuff correctly and
try again.
Er4 Pneumatic system leakage or cuff is
too loose during inflation
Er5 Inflation above 300mmHg
Measure again after 5
minutes. If the monitor is still
abnormal, please contact the
local distributor or the
factory.
Er6 Over 3 minutes while pressure
above 15 mmHg
Er7 EEPROM accessing error
Er8 Device parameter check error
Er9 MCU self-verity error
ErA Span error: pressure sensor
parameter error
DIA
mmHg
SYS
mmHg
PUL/min
9
M
8
9
M
8
9
7
114
TOPCOM BPM Arm 5500 XLD
3 Krevní tlak
3.1 Co je to krevní tlak?
Krevní tlak je tlak vyvíjený na stnu tepny, když protéká tepnami krev. Tlak namený pi
srdení kontrakci, když je krev vytlaována ze srdce, se nazývá systolický (nejvyšší).
Tlak namený pi rozšiování srdce, když se krev znovu do srdce nalévá, se nazývá
diastolický (nejnižší) krevní tlak.
3.2 Pro se má mit krevní tlak?
Mezi rznými problémy, kterými trpí moderní populace, jsou problémy spojované s
vysokým krevním tlakem zdaleka nejbžnjší. Nebezpen silná korelace vysokého
krevního tlaku a kardiovaskulárních onemocnní a vysoké úmrtnosti znamená, že
mení krevního tlaku je pro stanovení ohrožených osob nezbytné.
3.3 Standard krevního tlaku
V rámci Svtové zdravotnické
organizace (WHO) a národního
programu pro osvtu o krevním tlaku byl
vypracování standard krevního tlaku,
podle kterého jsou stanoveny oblasti
krevního tlaku s nízkým a vysokým
rizikem. Standard však má pouze
obecnou platnost, nebo krevní tlak
jednotlivých osob se mezi rznými lidmi,
rznými skupinami atd. liší.
Indikátor klasifikace SZO na pístroji
BPM Arm 5500 XLD bude spolen s
výsledkem ukazovat tídu krevního tlaku.
Je velmi dležité, abyste pravideln navštvovali svého lékae. Váš léka vám ekne,
jaký je váš normální rozsah krevního tlaku, stejn jako jaký bod pro vás pedstavuje
riziko.
Klasifikace krevního tlaku Systolický
(mmHg)
Diastolický
(mmHg)
Barevné
oznaení
Fáze 3 - hypertenze >180 >110 ervená
Fáze 2 - hypertenze 160-179 100-109 Oranžová
Fáze 1 - hypertenze 140-159 90-99 Žlutá
Vyšší - Normální 130-139 85-89 Zelená
Normální 120-129 80-84 Zelená
Optimální <120 <80 Zelená
Systolick (mmHG)
Diastolick (mmHg)
Mírná hypertenze
Slabá hypertenze
Vyšší - Normální
Normální
Závažná hypertenze
12
6
5
4
3
2
1
114


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Topcom bpm 5500 xld at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Topcom bpm 5500 xld in the language / languages: English, German, Dutch, Danish, French, Italian, Polish, Portuguese, Swedish, Turkish, Spanish, Norwegian, Finnish as an attachment in your email.

The manual is 1,82 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info