Page 11.13
11.10. LEAKAGE CURRENT PROBLEMS WITH HITACHI E5
INVERTER
The introduction of Hitachi inverter version E5 on the machine has given rise to problems with
increased earth leakage currents, which result in non-conformance to the relevant regulations for
electromedical equipment.
In fact, in order to keep the emissions within the limits set down by the regulations for light
industry, this inverter version has a capacitive filter to ground on the supply conductors, which
increases the current in the earth conductor up to 1200 µA.
The leakage currents for the individual machine components are detailed in the table below:
COMPONENT Model
I
leakage
Inverter Hitachi J100 E5
1200 µA
Power supply Lambda ZT30-522
300 µA
Schaffner FN350-20-29
4900 µA
Filter
Schaffner FS5458-20-29
350 µA
Transformer - -
with FN350 filter
5400 µ
µµ
µA
TOTAL FOR MACHINE
with FS5458 filter
1850µ
µµ
µA
To ensure conformance to the values prescribed by the regulations in force, pending further
instructions proceed as follows:
On the Lambda ZT30-522 power supply, eliminate the two capacitors C3 and C4 (2200 pF).
This reduces the leakage of the power supply to approximately 0 µA.
On the Hitachi E5 inverter, eliminate the two capacitors CNX and CPX (10 nF). This reduces
the leakage of the power supply to approximately 0 µA.
In place of the Schaffner 350/20/29 filter, use the Schaffner FS5458-20-29 filter specially
manufactured for Technogym, which has order code 0WB018. This filter is installed as standard
on machines with SN 6000120008, or raise the filter off the ground.
Suspend the Schaffner 350-20-29 filter using insulating spacers and plastic screws. This reduces
the leakage of the filter to approximately 0 µA.
Replace the Schaffner 660-3/06 filter with the Schaffner 610B-3/06 filter with order code
0WB016.
Replace the Schaffner 660-10/06 filter with the Schaffner 610B-10/06 filter with order code
0WB015.
Always obtain authorization from the Technogym Service Center before carrying out
this operation.
With these modifications, the machine leakage current is reduced to approximately 400 µA,
bringing it within the 500 µA limit prescribed by the regulations.
Downloaded from www.Manualslib.com manuals search engine
179


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Technogym Runrace at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Technogym Runrace in the language / languages: English as an attachment in your email.

The manual is 9,26 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info