Switch on Wheelchair scale
First, switch on the wheelchair scale.
Send the mode changing command (M) from the personal computer.
When necessary
(cautions for safety)
31
When necessary
(cautions for safety)
30
PC mode (continued)
1
Set tare weight
The tare weight need not be set if not required. It will then be treated as 0.
Ex
: 30 kg tare weight
PC
ee
PW-630
If commands accepted
PW-630
ee
PC
If settings incorrect
PW-630
ee
PC
LF
CR6E
LFCR0.03,tP,0D
LFCR0.0300D
2
Set height
Ex
: 171 cm
PC
ee
PW-630
If commands accepted
PW-630
ee
PC
If settings incorrect
PW-630
ee
PC
LF
CR6E
LFCR171,mH,3D
LFCR171003D
3
Set ID
Ex
: 0000012345
PC
ee
PW-630
If commands accepted
PW-630
ee
PC
If settings incorrect
PW-630
ee
PC
LF
CR6E
LFCR”5432100000”DI,5D
LFCR5432100000D
4
Confirm settings
PC
ee
PW-630
If setting is completed
PW-630
e
e
PC
If setting is incomplete
PW-630
e
e
PC
LF
CR”000000
0000”DI,5D,0,mH,3D,0.0,tP,0D
LFCR”543210000
0”DI,5D,171,mH,3D,0.03,tP,0D
LFCR?D
5
Example of Use
The wheelchair scale is used remotely as follows.
Start measurement (BMI mode)
PC
ee
PW-630
After measurement preparations is completed
PW-630
ee
PC
After measurement is completed
PW-630
e
(measurement result)
e
PC
After a person of measurement went down from Platform
PW-630
ee
PC
LF
CR1S
LFCR
LFCR6S
LFCRG
6
Measurement result
Readings will be outputted when measurement is finished.
If concluded normally
7
Next measurement
After the measurement is completed, LCD of PW-630 turns into a setting confirmation screen.
Data except for clothes weight are cleared.
8
Measurement cancel
PC
ee
PW-630
LF
CRq
9
Error
Scale error
PW-630
ee
PC
Measured value error
PW-630
ee
PC
LF
CR4E
LFCR0E
10