25
How to Use
24
How to Use
Connecting to Your Computer
1.Slide the
impedance
meter down
from the top.
2.
Hang the
switch hanger
on the main
unit hook
Specifications
Communication standard
Communication method
Signal speed
Data bit length
Parity
Stop bit
Flow control
Terminator
EIA RS-232C compatible
Asynchronous transaction
9600 bps
8 bits
None
1 bit
None
CR+LF
For Your Reference
A RS232C connector, D-sub
9 pin (female), is supplied
as standard with the main
unit.
Cables are not supplied and
need to be purchased
separately.
RS232C: straight D-sub 9 pin
(male) – D-sub 9 pin (female)
*Max. 3m
Data
Because data is output immediately following measurement regardless of the status of
the receiving side (personal computer, etc.) it is necessary to ensure that the receiving
side is always in receiving mode prior to conducting measurements.
(1) Output data format
Measurement data is output in the following format.
•Data segments are divided by commas (,).
•Terminators (the end of data) are CR (ASCll code 0DH), LF (ASCll code 0AH).
Model No.
MO
Gender Trunk % fat
Estimated Waist
Visceral fat level
Check Sum
XXXXXX GE 1 or 2 FT XX.X WC XXX.X IF XX.X CS XX
(2) Output Data Items
Item Format Content
Output Order
Standard
Over 130cm
Control data {0 Fixed at 16 1-5 byte variable length 1 1
Control data
˜0 Fixed at 1 1-5 byte variable length 2 2
Control data ˜1 Fixed at 1 1-5 byte variable length 3 3
Control data ˜2 Fixed at 1 1-5 byte variable length 4 4
Model number MO “AB-140” 8 byte fixed length 5 5
Gender GE 1 or 2 1 byte fixed length (1: Male, 2: Female) 6 6
Trunk % fat FT XX.X
3-4 byte variable length, 1 decimal place (unit: %)
77
Waist circumference
WC XXX.X
4-5 byte variable length, 1 decimal place (unit: cm)
8 X
Visceral fat level
IF XX.X 3-4 byte variable length, 1 decimal place 9 8
Check sum CS XX 2 byte fixed length 10 9
Header
7
Storing the impedance meter
Always store the impedance meter in the main unit upon completion of
measuring (the impedance meter transfer mode will end and the
impedance meter will return to a state ready for use).
8
End measuring
Press the key on the main unit to turn off.
*If the main unit power is not turned off, it will automatically turn off after the
results have been displayed for approximately two minutes.
Check that the
Power LED is off
For Your Reference
Take care not to catch the body hair or skin of
the subject when moving the impedance meter
on the abdomen. Use both hands on either side
of the impedance meter, as shown in the
diagram.
Measuring
waist circumferences over 130cm (cont’d)