Guidance and Manufacturer’s Declaration Electromagnetic Immunity
Radiated RF EM fields IEC
61000-4-3
All device input and
output connections
and cables.
3V/m
80 MHz to 2700MHz
80% AM at 1kHz
Radiated electromagnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Radiated RF Wireless
communication equipment
IEC 61000-4-3
All device input and
output connections
and cables.
See Table A below
This device has been subjected to
RF wireless communication bands
from cell phones, and other
communication devices
Electrical fast
transient/burst IEC 61000-
4-3
All device input and
output connections
and cables.
± 2kV for power supply
lines
100kHz repetition
frequency Mains power quality should be
that of a typical commercial or
hospital environment
(Professional Healthcare Facility)
Surge IEC 61000-4-5
All device input and
output connections
and cables.
± 0.5, 1, 2kV
All device input and
output connections
and cables.
± 0.5, 1kV
Conducted Disturbances
induced by RF fields IEC
61000-4-6
All device input and
output connections
and cables.
3V
0.15MHz 80MHz
6V in ISM bands
between 0,15 MHz and
80MHz
80% AM at 1kHz
Mains power quality should be
that of a typical commercial or
hospital environment. All handheld
and patient coupled parts should
be consistent with intended use.
DC Input and all
cables (>3m)
Power Frequency magnetic
field IEC 61000-4-8
All device input and
output connections
and cables.
30 A/m
50 or 60 Hz
Power Frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: a) U
T
is the a.c. mains voltage prior to application of the test level
b) E.g. 25/30 means 25 periods at 50 Hz or 30 periods at 60 Hz
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Device input (AC
power)
0% U
T
: 0.5 cycle
a)
At 0, 45, 90⁰, 135⁰,
180⁰, 225⁰, 270⁰ and
315⁰
Mains power quality should be
that of a typical commercial or
hospital environment
(Professional Healthcare Facility)
If the user of the monitor requires
continued operation during power
mains interruption, it is
0% UT: 1 cycle
70% UT: 25/30 cycles b)
Single Phase: at 0
46


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for SunTech CT40 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of SunTech CT40 in the language / languages: English as an attachment in your email.

The manual is 2,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of SunTech CT40

SunTech CT40 User Manual - German - 49 pages

SunTech CT40 User Manual - Dutch - 52 pages

SunTech CT40 User Manual - French - 50 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info