18
INTRODUCTION
Soehnle. Otherwise, danger may
arise for the user and the device.
Please inform Soehnle customer
service of unexpected operating
status or events.
The adapter is part of the ELEC-
TRONIC MEDICAL DEVICE.
The plugs/pins of the adapter
plug isolate the device against the
mains power supply.
Please do not place the device in
a location in which it is difficult to
disconnect from the power supply.
Keep the device outside of the
reach of small children/pets to
prevent the inhalation or swallo-
wing of small parts. The cable/cord
poses a risk of strangulation.
When stored at the minimum sto-
rage temperature between uses,
electrical medical devices must be
warmed up for at least 30 minutes
before they are operational. When
stored at the maximum permissib-
le storage temperature between
uses, electrical medical devices
must cool off for at least 30 minu-
tes before they are operational.
Please carefully monitor a possible
interruption of blood circulation
and a possible injury of the patient.
Do not place the cuff over a
wound as this may cause further
injuries.
Do not place the cuff on the same
arm on which other electrical
medical devices are placed.
Please ensure that the operation
of the device does not interrupt
blood circulation for a longer
period.
Avoid squeezing or constriction of
the connecting tube.
The Blood Pressure Monitor has
been tested pursuant to the requi-
rements of EN 1060-4:2004.
Metrological maintenance and care
information
Do not perform any metrological
maintenance if the device is in use.
For reliable operation, we recom-
mend that the device is subject to
metrological control once every
two years pursuant to Sec. 14
of the Medical Product Opera-
ting Directive (MPBetriebV). We
recommend testing the devices
performance once every two years
and after metrological control and
repair, at a minimum by testing
the tolerance requirements of the
pressure display and tightness
of the cuff (testing at least at 50
mmHg and 200 mmHg).
The metrological control must be
performed by a licensed tester or
testing institute.
We recommend cleaning the de-
vice and cuff only with a soft cloth.
Storage information
If the device is not being used,
store it and the adapter in a dry lo-
cation away from heat, dust, fibres,
lint, and out of direct sunlight. Do
not place heavy objects on the
storage box.
Information about handling batteries
Do not mix new and used bat-
teries.
20


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Soehnle 68108 - Systo Monitor 200 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Soehnle 68108 - Systo Monitor 200 in the language / languages: English, German, Dutch, French, Italian, Polish, Swedish, Spanish as an attachment in your email.

The manual is 1,37 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info