Electrical connections
The power supply has to be connected to +~ and . is considered as system
ground. The same ground reference has to be used for the eSENSE unit and for the
DDC/signal receiver.
PLEASE NOTE! The same ground reference has to be
used for the eSENSE unit and for the control system!
Terminal
Function
Electrical data
Remarks
+~
Power (+)
Power ground
(-)
24 VAC/DC+ (+-
20%), 2W
24 VAC/DC-
System voltage reference
OUT1
OUT2
Analogue
output 1 (+)
Silences the
acoustic alarm
0-10 VDC
0-2000 ppm CO
2
A push on the push button
silences the acoustic alarm for 30
minutes.
Table I. Connections of the main terminal of eSENSE
LED Colour
Electrical data
Remarks
Green
0VDC
10VDC
Lit between 0-800 ppm CO
2
Yellow
0VDC
10VDC
Lit between 800-1400 ppm CO
2
Red
0VDC
10VDC
Lit above 1400 ppm CO
2
. Buzzer sounds.
Table II. The LEDs
Self-diagnostics
The system contains complete self-diagnostic procedures that are executed
automatically when the sensor is in operation. Sensors with display show a wrench if
an error is found. The wrench is shown during the first seconds after power up and if
the measuring range is exceeded.
Maintenance
The eSENSE is basically maintenance free in normal environments thanks to the
built-in self-correcting ABC algorithm. Discuss your application with your distributor in
order to get advice for a proper calibration strategy.
PLEASE NOTE! The sensor accuracy is defined at continuous operation (at
least 3 weeks after installation)
Electronic products should be disposed of via a suitable recycling centre.
3


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Senseair eSENSE II FAI Disp at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Senseair eSENSE II FAI Disp in the language / languages: English as an attachment in your email.

The manual is 0,25 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info