528660
15
Zoom out
Zoom in
Previous page
1/24
Next page
11
2.Lumidisc
Employ for illumination contrast adjustment and illuminance measure-
ment.
6. STANDARD ACCESSORY OPERATION
Main light
Fill-in light
Contrast measurement
a. Mount Lumidisc @3 to light
sensor w. Install in the same
manner as Lumisphere. (See
P. 8)
b. Hold meter at subject position
and point Lumidisc @3 directly
toward center of main light.
c. Press stopper button !3 and
read meter scale footcandle
indication e.
d. Turn the memo pointer dial
and set the memo pointer !7
at the footcandle value the
needle points to.
e. In the same manner, measure
the fill-in light. In this case,
take care that light from main
source does not enter sensor.
f. Read a difference between
the footcandle value the
memo pointer is set (see
above d) and the one for fill-
in light (= needle position).
Ratio between main and fill-in light footcandles becomes the con-
trast ratio.
Example:With main light at 640 footcandles, fill-in light reads 320 foot-
candles. Illumination contrast ratio becomes;
640 2
320 1
or 2 : 1
=
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sekonic L-398A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sekonic L-398A in the language / languages: English as an attachment in your email.

The manual is 1,14 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Sekonic L-398A

Sekonic L-398A User Manual - German - 24 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info