50947
11
Zoom out
Zoom in
Previous page
1/13
Next page
page 10
Manual SA 230
1 The “program” mode is made accessible by pressing key during
switching on the scale with the on/off switch.
As soon as the display shows : CAL release the key.
2 Press one after the other the following key’s:
The display shows now: rA 1 of : rA 0. Select with the key:
1 = 0,5% tolerance check on “quick calibration” is in operation
0 = 0,5% tolerance check on “quick calibration” is disabled
3 Press key : . The internal measuring value of the empty platform
is displayed. This value should be between 10.000 en 40.000 .
4 Press key : as soon as the reading has come to a standstill.
The zero value is entered now.
5 Put on the platform an accurate calibration weight. See table below.
The available values in this program are dependent of the weighing
capacity of the scale.
Weighing capacity Choice of the
type of scale calibration weight
300 g 100, 200
, 300 g
600 g 200, 400, 500
, 600 g
1200 g 500, 1000
, 1200 g
1550 g 500, 1000
, 1500 g
3100 g 1000, 2000
, 3000 g
The figures in this table indicated as fat
will realise the best accuracy for
that scale, in relation to the full weighing range.
Make use of weights with sufficient accuracy, at least class M1.
6 Press key : as soon as the reading has come to a standstill.
Now the available calibration values are show one after another on the
display.
7 Press again key if the correct value is displayed.
After a few seconds the calibration value is calculated and entered.
The calibration procedure will terminate now automatically.
The scale switches back to the weighing mode.
Switch the scale off and on again.
WEEGTECHNIEK HOLLAND B.V. Website www.satex.nl
Tel + 31 36 522 20 30
Patroonsweg 23-27 Email info@satex.nl Fax + 31 36 522 20 60
3892 DA ZEEWOLDE The Netherlands
Complete Weight Calibration
NET MD
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Satex SA 230 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Satex SA 230 in the language / languages: English as an attachment in your email.

The manual is 0,72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Satex SA 230

Satex SA 230 User Manual - Dutch - 13 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info