Guidance and manufacturer’s declaration – electromagnetic immunity
The SFT77 Multi-function Thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the SFT77 Multi-function Thermometer should assure that it is used in such an environment.
IMMUNITY test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floor should be wood, concrete or ceramic tile. If floors
are covered with shythetic material, the relative humidity
should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV air for input/out
-
put lines
Not applicable
Mains power quality should be that of a typical com
-
mercial or hospital environment.
Surge
IEC 61000-4-5
±1kV differential mode
±2kV common mode
Not applicable
Mains power quality should be that of a typical com
-
mercial or hospital environment.
Voltage dips, short interrup
-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
for 0.5 cycle
40% U
T
(60% dip in
U
T
for 5 cycles
70% U
T
(30% dip in
U
T
for 25 cycles
5% U
T
(>95% dip in
U
T
for 5 sec
Not applicable
Mains power quality should be that of a typical com
-
mercial or hospital environment. If the use of the SFT77
Multi-function Thermometer requires continued ope
-
ration during power mains interruptions, it is recom-
mended that the SFT77 Multi-function Thermometer
be powered from an uninterruptible power supply or
a battery.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer
-
cial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Table 2
67


Need help? Post your question in this forum.

Forumrules
1

Forum

sanitas-sft-77

Reset search

  • After changing batteries CLr came on screen and it doesn’t work
    Submitted on 5-3-2023 at 22:02

    Reply Report abuse
  • Changed batteries. Screen shows C(celcius) symbol for couple of seconds. After that OFF and turns off. Can't be turned on again. Broken? Submitted on 7-5-2021 at 10:10

    Reply Report abuse
  • After I chaged batteries, the screen only show ”CLr” and i can’t use thermometer anymore, is there something what I can do to make it work Submitted on 5-5-2021 at 19:30

    Reply Report abuse
  • After I chaged batteries, the screen only show ”CLr” and i can’t use thermometer anymore, is there something what I can do to make it work or do I have to buy new? Submitted on 24-8-2020 at 20:50

    Reply Report abuse
    • I had the same. Clear sensor to perfect clean. Remove batteries, wait couple minutes and insert batteries with "head" button preset. I make couple attemps and its work again ;) Answered on 15-3-2021 at 22:26

      Vote up (23) Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sanitas SFT 77 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sanitas SFT 77 in the language / languages: Dutch, French, Portuguese as an attachment in your email.

The manual is 5,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info