579873
5
Zoom out
Zoom in
Previous page
1/56
Next page
5
If pregnant
In case of cancer
After operations where increased muscle contractions might impair the
healing process
Never use near the heart. Stimulation electrodes should
never be placed anywhere on the front of the thorax
(marked by ribs and breastbone), but above all not on the
two large pectoral muscles. Here it can increase the risk of
ventricular fibrillation and lead to cardiac arrest
On the bony part of the cranium, near the mouth, throat or larynx
Near the throat / carotid artery
Near the genitals
On acutely or chronically diseased (injured or inflamed) skin (e.g. in the
event of painful and painless inflammation, reddening, skin rashes (e.g.
allergies), burns, bruises, swellings, open wounds and wounds in the
process of healing, on operation scars in the process of healing)
In environments with high humidity such as in the bathroom or when
having a bath or shower
Do not use after consuming alcohol
When a high frequency surgical device is connected at the same time
Before using the machine, you should consult with the doctor in charge
of your treatment in the event of the following:
Acute diseases, especially if hypertension is suspected or actually exists,
or blood-clotting disorders, tendency to thromboembolic diseases or in
case of malignant new masses
All skin diseases
Chronic pain disorders which have not been clarified, regardless of the re-
gion of the body
Diabetes
All sensitivity disorders with reduced pain awareness (such as metabolic
disturbances)
Medical treatments carried out simultaneously
Complaints arising as a result of the stimulation treatment
Constant skin irritation as a result of prolonged stimulation at the same
electrode point.
IMPORTANT!
Use the Digital EMS/TENS solely:
On humans
For the purpose for which it was developed and in the manner specified in
these instructions for use. Any improper use can be hazardous.
For external application.
With the original accessory parts which are supplied and can be re-or-
dered, otherwise the warranty becomes null and void.
PRECAUTIONARY MEASURES:
Always remove the electrodes from the skin with a moderate pull in order
to avoid injury in the event of highly sensitive skin.
Keep the machine away from heat sources and never use it near (~1 m)
short-wave or microwave equipment (e.g. mobile phones), as this may
lead to unpleasant current spikes.
Do not expose the machine to direct sunlight or high temperatures.
Protect the machine from dust, dirt and humidity. Never submerge the de-
vice in water or other liquids.
The machine is suitable for private use.
For reasons of hygiene, the electrodes should only be used on one person.
If the machine is not functioning properly, or if you start feeling unwell or
there is any pain, stop the application immediately.
To remove or move the electrodes, first switch off the machine or the ap-
propriate channel in order to avoid unwanted irritation.
Never modify any electrodes (e.g. by cutting). This leads to a higher
current density and can be dangerous (max. recommended output value
for electrodes: 9 mA/cm², an effective current density over 2 mA/cm² re-
quires greater attention).
Do not use when asleep, when driving a vehicle or at the same time as op-
erating machinery.
Never use with any activities where an unforeseen reaction (e.g. increased
muscle contraction despite low intensity) may be dangerous.
Make sure that no metallic objects such as belt buckles or necklaces can
come in contact with the electrodes during stimulation. If you wear any
jewellery or piercings (e.g. belly piercing) near the application site, you
must remove these before using the machine as burning may otherwise
occur at some points.
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sanitas SEM 40 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sanitas SEM 40 in the language / languages: English, Danish, Swedish, Finnish as an attachment in your email.

The manual is 3,79 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Sanitas SEM 40

Sanitas SEM 40 User Manual - German, French, Italian, Portuguese - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info