Measurement range Cuff pressure 0-300 mmHg,
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display accuracy Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 74 mm x W 72 mm x H 24.5 mm
Weight Approximately 100 g (without bat-
teries)
Cuff size
135 to 195 mm
Permissible operating
conditions
+5°C to +40°C, 15-93% relative humi-
dity (non-condensing)
Permissible storage
conditions
-25°C to +70°C, ≤ 93% relative humi-
dity, 700-1050 hPa ambient pressure
Power supply 2 x 1.5V
AAA batteries
Battery life For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Accessories Instructions for use, 2 x 1.5V AAA
batteries, storage box
Classification Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
This device complies with European Standard EN60601-1-2
and is subject to particular precautions with regard to electro
-
magnetic compatibility. Please note that portable and mobile
HF communication systems may interfere with this unit. More
details can be requested from the stated Customer Services
address or found at the end of the instructions for use.
The device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinpro-
duktgesetz) and the standards EN1060-1 (Non-invasive sphy-
gmomanometers – Part 1: General requirements), EN1060-3
(Non-invasive sphygmomanometers – Part 3: Supplemen
-
tary requirements for electro-mechanical blood pressure
measuring systems) and IEC80601-2-30 (Medical electrical
equipment – Part 2-30: Particular requirements for the basic
safety and essential performance of automated non-invasive
sphygmomanometers).
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sanitas SBC 22 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sanitas SBC 22 in the language / languages: English, German, French, Italian as an attachment in your email.

The manual is 9,73 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info