the systolic or diastolic pressure was outside the measure-
ment range (
Hi
or
Lo
appears on the display)
the cuff is fastened too tightly or too loosely
(
or
appears on the display)
the blood pressure is higher than 297 mmHg for more than
1.5 seconds (
appears on the display)
pumping up takes longer than 180 seconds (
appears on
the display)
there is a system or unit error (
, , or
appears
on the display)
the batteries are almost empty
8. Cleaning and storing the device and cuff
Clean the device and cu carefully using a slightly damp cloth
only.
Do not use any cleaning agents or solvents.
Under no circumstances hold the device and cu under wa-
ter, as this can cause liquid to enter and damage the device
and cu.
If you store the device and cu, do not place heavy objects on
the device and cu. Remove the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical specifica-
tions”, the technical alarm will appear on the display indicat-
ing either “
  ” or “ Lo . In such cases, you should seek medical
assistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.
9. Technical specifications
Model No. BC 28
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the wrist
Measurement range Cuff pressure 0 – 300mmHg,
Systolic 60-260mmHg,
Diastolic 40-199mmHg,
Pulse 40 –180 beats/minute
Display accuracy Systolic ± 3mmHg,
Diastolic ± 3mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 83.8mmx W59.9mmx H29.4mm
Weight Approximately 92g (without batteries)
Cuff size 140 to 195mm
Permissible operating
conditions
+ 10°C to +40°C, ≤85% relative humid-
ity (non-condensing)
Permissible storage
conditions
-20°C to +50°C, ≤ 85% relative humid-
ity, 700 –1060hPa ambient pressure
Power supply
2 x 1.5V
AAA batteries
23
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sanitas SBC 15 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sanitas SBC 15 in the language / languages: English, German, French, Italian as an attachment in your email.

The manual is 8,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info