Introduction
Error code for your reference
Rossmax Fingertip Pulse Oximeter SB100 is used
to measure arterial oxygen saturation (% SpO2) of
hemoglobin and pulse rate, an important indicator
of your respiratory function. It is non-invasive device
intended for spot-check of adult and pediatric whose
age is over 3 at home, hospital.
Attention: Consult the accompanying
documents. Please read this manual carefully
before use. Please be sure to keep this manual.
SENSOR ERROR:
Sensor cannot be detected,
return the device to your
local distributor or service
centre.
MEASURE ERROR:
Signals cannot be detected,
turn the device off and
measure again.
Installing Batteries
1. Use thumb to slide battery
cover out
2. Insert or replace 2 “AAA” sized
batteries down with the
correct electrical polarity.
You need to replace the
batteries when
1. Battery icon is blinking on
display.
2. The function button is pressed
and nothing appears on
display.
Caution: Batteries may leak or explode if used or
disposed of improperly. Remove batteries if
the device will be stored for long time. Do
not use different types or brands of batteries
at the same time. Do not mix fully charged
and partially charged batteries at the same
time.
Attaching the lanyard
1. Insert the narrow end of
the lanyard through the
holder.
2. Draw the other end of the
lanyard through the loop
at the narrow end and
tighten.
How to use
1. Open the clip; press the Power On button as .
2. Information of software version appears; insert one
finger, nail side up, into the finger opening of the
pulse oximeter.
Note: if no finger insert, the device will auto shut off
after 8 seconds.
3. The pulse strength indication shows , pulse
oximeter begins its measurement as .
Note: make sure the finger is lying flat, Do not shake
and keep body steady during measurement as
.
4. Your SpO2 and pulse rate values will appear on the
screen after few seconds as .
Note: 1. Don’t remove your finger until the measurement
is completed.
2. If SpO2 and pulse rate cannot be detected,
“ will appear on the screen as .
3. While pulse strength is low, the reading will
flicker.
Note:
1. The SpO2 sensor and photoelectric receiving tube
should be arranged in a way with the subjects
arteriole in a position there between.
2. Make sure the optical path is free from any optical
obstacles like rubberized fabric.
3. Excessive ambient light may affect the measuring
result. It includes fluorescent lamp, dual ruby light,
infrared heater, direct sunlight and etc.
4. Strenuous action of the subject or extreme
electrosurgical interference may also affect the
accuracy.
Cautionary Notes
- This device is to be operated by trained personnel
only.
- This device has no audible and it intended only for
spot-checking, but not medical result evaluation.
- This device is designed to determine the percent-
age of arterial oxygen saturation of functional hemo-
globin. Factors that may degrade pulse oximeter per-
formance or aect the accuracy of the measurement
include the following:
Do not apply the pulse oximeter on the same arm
as a blood pressure cu, arterial catheter or infusion
line(s)
Excessive light, such as sunlight or direct home light-
ing.
Not steady at the site of application (e.g. trembling)
Moisture in the device
Improperly applied device
Finger is too large or too small to t into the device.
Poor pulse quality
Venous pulsations
Anemia or low hemoglobin concentrations.
Cardiogreen and other intravascular dyes
Carboxyhemoglobin
Methemoglobin
Dysfunctional hemoglobin
Articial nails or ngernail polish
On ngers with anatomical changes, oedemas, scars
or burns.
- Using the device for long periods may cause pain
for people with circulatory disorders. Reposition the
device at least once every 4 hours to allow the pa-
tient’s skin to breath and to check patient’s condition
regularly.
- Do not use the device near ammable or explosive
gas mixtures.
- Do not use the device during an MRI or CT scan, be
used no closer than 30 cm (12 inches) to any part
of the [ME EQUIPMENT or ME SYSTEM], including ca-
bles specied by the manufacturer
- The device may not work when circulation is re-
duced. Warm or rub the nger, or re-position the
device.
- This device is a precision electronic instrument and
must be repaired by qualied technical profession-
als. Field repair of the device is not possible. Do no
attempt to open the case or repair the electronics.
Opening the case may damage the device and void
the warranty.
- Do not overextend the device’s spring.
- A functional tester cannot be used to access the ac-
curacy of a pulse oximeter monitor.
- Do not self-diagnose or self-medicate on the basis of
the measurements without consulting your doctor.
In particular, do not start taking any new medication
or change the type and/or dosage of any existing
medication without prior approval.
- Do not look directly inside the housing during the
measurement. The red light and the invisible infra-
red light in the pulse oximeter are harmful to your
eyes.
- This device is not intended for use by people (in-
cluding children) with restricted physical, sensory or
mental skills or a lack of experience and/or a lack of
knowledge, unless they are supervised by a person
who has responsibility for their safety or they receive
instructions from this person on how to use the de-
vice. Children should be supervised around the de-
vice to ensure they do not play with it.
- Neither of the displays for the pulse wave and pulse
bar allows the strength of the pulse or circulation to
1. SpO2 icon
2. Pulse strength
indication
3. Pulse rate icon
7
1
2
3
4
5
6
4. Beats per minute
5. Power On Button
6. Battery icon
7. Battery compartment
Name/ Functions of each part
2


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Rossmax SB100 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Rossmax SB100 in the language / languages: English as an attachment in your email.

The manual is 1.96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Rossmax SB100

Rossmax SB100 User Manual - German - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info