Rossmax Neb Tester is a portable and easy-to-use device that allows the user to quickly
check the performance of the pump of a compressor nebulizer. Power source is not
required for Neb Tester during operation and the test result is easy to read and interpret.
Neb Tester consists of an oil pressure gauge, ow meter, air tube and stainless steel stand.
This device is used to simulate the working conditions of the pump when the nebulizer is
in operation with a medication cup, and you are able to check:
1. Maximum (free) air ow
2. Operational air ow at a particular pressure
3. Maximum air pressure
Parts name/location
Introduction
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
Flow meter
Regulator
Pressure
gauge
Air tube
Insert air tube into connectorPlace the device in a
horizontal position
Open the rubber vent-plug
and then close it after
Back view
Installation
Operation procedures
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
1. Check maximum (free) air ow
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Adjust air ow and set the level of the pressure at 0 PSI by turning the
regulator of ow meter.
Step 6: Check the air ow rate.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
2. Check operational air ow at a particular pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the pressure to a desired level by adjusting the regulator of the ow
meter.
Step 6: Check the air ow rate.
Pressure (PSI) Flow rate (L/min)
NA100 / NB500 (N1) 15.0 ≥ 3.5
NE100 / NF100/ NJ100 (N4) 15.0 ≥ 3.5
NK1000/ NB80 / NF80 (N1) 8.0 ≥ 3.0
NB60 / NI60/ NH60 / NF60 (N1) 8.0 ≥ 2.0
NL100 (N2) 10.5 ≥ 3.5
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
3. Check maximum air pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the air ow rate at 0 L/min by turning the regulator of ow meter
clockwise completely.
Step 6: Check the values on the pressure gauge.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
Air
compressor
connector
Rubber
vent-plug
2


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Rossmax Neb Tester at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Rossmax Neb Tester in the language / languages: English as an attachment in your email.

The manual is 1.92 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info