651365
10
Zoom out
Zoom in
Previous page
1/12
Next page
Q
ACNE
Advanced blue light
acne clearing system
9
MANUFACTURER’S DECLARATION OF EMC
Manufacturer’s declaration – electromagnetic immunity
The LITE3 is intended for use in the electromagnetic environment specified below.
Immunity Test
IEC 60601
test level
Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) IEC 61000-
4-2:2009
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Conducted RF
IEC 61000-4-6:2009
Radiated RF immunity
IEC 61000-4-3:2006
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
1Vrms
3V/m
80MHz to 2.5GHz
Portable and mobile RF communications equipment should be
used no closer to any part of the LITE 2, including cables, than
the recommended separation distance calculated from the equa-
tion applicable to the frequency of the transmitter.
Recommended separation distance
d=3.5ÖP
d=1.2ÖP 80MHz to 800MHz d=2.3ÖP 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Electrical fast transient/burst IEC 61000-
4.4:2004 +A1
± 2 kV for power supply lines ± 2 kV for power supply lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5:2005
± 1kV line(s) to line(s)
± 2kV line(s) to earth
± 1kV differential mode
± 2kV common mode
Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short interruptions and volt-
age variations on power supply input lines
IEC61000-4-11:2004
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60% dip in UT) for 5 cycles
70 % UT
(30% dip in UT) for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
<5 % UT
(>95 % dip in UT) for 0,5 cycle
40 % UT
(60% dip in UT) for 5 cycles
70 % UT
(30% dip in UT) for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
Mains power quality should be that of a typical commercial or
hospital environment.
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8:2010
3 A/m 3 A/m
Note 1: UT is the a.c. mains voltage prior to application of the test level. Note 2: Please refer to IEC60601-1 for further clarification
© The Dezac Group 2015
© The Dezac Group 2015
© The Dezac Group 2015
© The Dezac Group 2015
10


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Rio Q-Acne - Blue Light Treatment System at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Rio Q-Acne - Blue Light Treatment System in the language / languages: English as an attachment in your email.

The manual is 1,91 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info