The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including
any d.c. component, shall not deviate by more than ± 20 % when measured with a load
resistance within the range specified.
If confirmation is required that the Revitive works within its Essential Performance aer a
certain period of time, contact the manufacturer
Output Specifications for Transcutaneous Electrical Nerve Stimulation (TENS):
Waveform Biphasic
Shape Rectangular
Maximum Output Voltage: @500Ω 50 Vp
@2kΩ
115
Vp
@10kΩ
135
Vp
Maximum Output Current: @500Ω
100mA
@2kΩ
57.5 mA
@10kΩ
13.5 mA
Pulse Width: 200µS (30µS inter phase delay)
Frequency (Hz) 80-130 Hz
For multiphasic waveforms only: Symmetrical phases Yes
Phase Duration 100µS
Net Charge (per pulse @ 500Ω) 0μC
Maximum Phase Charge (@ 500Ω) 10μC
Maximum Current Density (@ 500Ω) 0.104 mA/cm²
Maximum Average Power Density (@ 500Ω) 4.19 mW/cm2
Technical Specications
38


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Revitive Medic at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Revitive Medic in the language / languages: English as an attachment in your email.

The manual is 4.63 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info