Pairing /Calibration
Calibrating the sensors (zero point calibration)
This product has a correction function
for the zero point uctuation caused
by varying temperatures. The
accuracy of this function improves by
calibrating the sensor when there is
a difference in temperature of more
than 4 °C.
This function cannot measure
correctly if you calibrate the sensor
before it is acclimated to the outside
temperature. The sensor requires
more than 20 minutes to become
acclimated.
Checking the Zero Point
1. Position the crank arm so
it is perpendicular to the
ground, pointing downward.
2. Tap [Pedaling Monitor R] in
the sensor list screen of the
SGX-CA500.
3. Conrm the value in [Force
Preview].
Make sure that the [Tangential Force]
and [Radial Force] values are as
shown here.
Tangential Force: 0 ± 3 N
Radial Force: 0 ± 3 N
Calibration of the right side is nished.
Calibrate the left side in the same way.
Calibration method in
power meter mode using a
cyclocomputer from another
manufacturer
For how to calibrate, refer to the User’s
Manual of the cyclocomputer you are
using.
27


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Pioneer SGY-PM910HL at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Pioneer SGY-PM910HL in the language / languages: English, German, Dutch, French, Italian, Spanish as an attachment in your email.

The manual is 4,01 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Pioneer SGY-PM910HL

Pioneer SGY-PM910HL Installation Guide - English - 26 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info