5.2 Settings
To get to the setup menu, press the ENTER key (6) until the first digit flashes. Then enter “888“.
With the Arrow right key (7), you can navigate through the digits and change the value of the digit
with the Arrow up key (8). Confirm with ENTER (6).
The following options can be selected using the Arrow up key (8):
Display
Meaning
Description
Ext
Exit
Back to normal measuring mode
S1
Pre-alarm
Enter the desired value (max. 50 m/s). You can move the
cursor with the Arrow right key (7) and change the value of the
digits with the Arrow up key (8). Confirm with ENTER (6).
Please note:
The pre-alarm value must not be higher than the alarm
value and the alarm value must not be lower than the pre-
alarm value.
S2
Alarm
Enter the desired value (max. 50 m/s). You can move the
cursor with the Arrow right key (7) and change the value of the
digits with the Arrow up key (8). Confirm with ENTER (6).
Please note:
The pre-alarm value must not be higher than the alarm
value and the alarm value must not be lower than the pre-
alarm value.
Flt
Filter
You can use the Arrow right key (7) to navigate through the
digits and the Arrow up key (8) to change the value of the
digits. Confirm with ENTER (6).
The following options can be selected:
000“ Current wind speed
Change interval of display: 200 ms
Change interval of relay: 200 ms
002“ 2-minute average value
Change interval of display: 120 s
Change interval of relay: 120 s
005“ 5-minute average value
Change interval of display: 300 s
Change interval of relay: 300 s
Str
Factory settings
Resets all parameters to factory settings
To enter the relevant menu, select the menu with the Arrow up key (8) and confirm with ENTER
(6).
You can leave the menu by selecting “Ext“ and confirming with the ENTER (6) key. If no key is
pressed for 60 seconds, the device enters normal measuring mode automatically.
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for PCE instruments PCE-WSAC 50 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of PCE instruments PCE-WSAC 50 in the language / languages: English, German, Dutch as an attachment in your email.

The manual is 0,78 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info