4. Error Messages and Troubleshooting
EN
27
4.2
Troubleshooting
In case of any of the below problems occur during measurement, first check that no
other electrical device is within 30cm. If the problem persists, please refer to the table
below.
Problem Cause Solution
The measurement result is extremely
high (or low).
Arm cuff is applied too loosely.
Apply the arm cuff tighter.
Refer to section 3.1.
Movement or talking during
measurement.
Remain still and do not talk during
measurement.
Refer to section 3.3.
Clothing is interfering with the arm
cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.1.
Arm cuff pressure does not rise.
The air connector is not securely
connected into the air jack.
Make sure that the air tube is
connected securely.
Refer to section 3.1.
Air is leaking from the arm cuff.
Replace the arm cuff with a new
one.
Refer to section 6.1.
Arm cuff deflates too soon. The arm cuff is loose.
Apply the arm cuff correctly so that
it is firmly wrapped around the arm.
Refer to section 3.1.
Cannot measure or the results are
too low or too high.
The arm cuff has not been inflated
sufficiently.
Inflate the arm cuff so that it is 30 to
40 mmHg above your previous
measurement result.
Refer to section 3.3.
31


Need help? Post your question in this forum.

Forumrules
1

Forum

omron-mit5s-connect----hem-7280t-e

Reset search

  • Model HEM-7280T cant export data as a pdf file as I have previously done Submitted on 1-10-2022 at 05:41

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron MIT5s Connect - HEM-7280T-E at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron MIT5s Connect - HEM-7280T-E in the language / languages: English as an attachment in your email.

The manual is 1,3 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron MIT5s Connect - HEM-7280T-E

Omron MIT5s Connect - HEM-7280T-E User Manual - German - 48 pages

Omron MIT5s Connect - HEM-7280T-E User Manual - Dutch - 48 pages

Omron MIT5s Connect - HEM-7280T-E User Manual - French - 48 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info