Abnormal measurement value
Cause / solution
The causes below are possible. Check the patient by palpation and then repeat measurement.
Body movement (chills or other trembling)
Arrhythmia.
Noise in the cuff
- A nearby person touched the patient.
- Cardiac massage was being performed.
The measurement value is questionable
Cause Solution
Deflates quickly Check for a loose cuff connection.
Simultaneously perform measurement with a stethoscope.
Place the stethoscope and listen while viewing the pressure display of the
manometer.
Blood pressure may vary widely due to physiological effects.
The causes below are possible.
Emotional excitement or agitation
- Pain due to cuff wrapping
- White coat hypertension
Cuff size or wrapping method not correct
Cuff wrapping position on upper arm not at the same height as the heart
Patient’s blood pressure not stable due to pulsus alternans, respiratory changes, or other reason
Incorrect cuff size used.
Measure circumference or patient’s arm and ensure
correct sized cuff is used.
Cuff wrapped over thick clothing.
Ensure cuff is applied to a bare arm, or very thin
clothing.
Patient not seated properly.
Ensure patient is seated, feet flat on the floor, cuff at
heart level.
Patient ate, drank, or exerted themselves recently.
Ensure before measurement taken, patient has not
had food, caffeinated or alcohol beverages, or
exerted/exercised in the last 30 minutes.
Stethoscope
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HBP-1300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HBP-1300 in the language / languages: English as an attachment in your email.

The manual is 0,53 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HBP-1300

Omron HBP-1300 User Manual - German - 38 pages

Omron HBP-1300 User Manual - Dutch - 38 pages

Omron HBP-1300 User Manual - French - 38 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info