Maintenance
Maintenance Inspection and Safety Management
The HBP-1300 must be maintained to ensure functionality and to secure the safety of patients and operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct
periodic inspections. We recommend that the verification test be performed at least once a year.
Cleaning of the Device
Cleaning and disinfecting should be performed in accordance with your facility’s infection control
practice.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).
If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Accessory Care
Non-Invasive Blood Pressure Measurement (NIBP)
Cuff
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.
Example of connection for pressure accuracy confirmation:
1. Display the pressure accuracy confirmation screen as explained in
“Utility Mode” on page 17.
2. Connect the blood pressure monitor, the calibrated reference pressure
gauge, and the cuff and inflation bulb.
3. Check the pressure value of the blood pressure monitor and the
pressure value of the calibrated reference pressure gauge.
Note:
Make sure that the blood pressure monitor reading is within ±3 mmHg compared to the calibrated
reference pressure gauge.
To rapidly deflate the cuff, press the [ ] button.
To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the procedure from
step 1 in “Utility Mode” on page 17.
Display the pressure value.
Display “0” in the pressure accuracy confirmation screen.
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HBP-1300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HBP-1300 in the language / languages: English as an attachment in your email.

The manual is 0,53 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HBP-1300

Omron HBP-1300 User Manual - German - 38 pages

Omron HBP-1300 User Manual - Dutch - 38 pages

Omron HBP-1300 User Manual - French - 38 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info