Failure, causes and rectification
Failure and their possible causes
Incorrect measurement is indicated by
1. A correct reading could not be obtained
because measurement was disturbed by
movement of the body.
2. The cuff is not fitted correctly.
3. Your clothing has constricted your
blood flow.
4. There is still air in the cuff when the
monitor is switched off.
The (START ) button was pressed
before the ( ) symbol was displayed.
The blood pressure values displayed are
ex tremely low or high, or they are
implausible.
The display does not light up when the
(START ) button is pressed.
1. The batteries are exhausted.
2. The +/- poles of the batteries have been
reversed.
3. The battery contacts are dirty.
The batteries are weak or exhausted.
The cuff pressure does not rise
although the pump motor can be heard.
problem with mem. function
E
Er
Rectification
1. Repeat the measurement keeping
perfectly still. Do not move your arm and
do not speak.
2. Check that the cuff is correctly fitted.
Then repeat the measurement.
3. Remove the item of clothing which
caused the constriction.
4. The unit may be defective. Please have
it examined and, if necessary, repaired
by OMRON.
Wait for the ( ) symbol indicating
readiness before you press the (START )
button.
Refer to the instructions and then repeat
the measurement.
1. Check the batteries and, if necessary, fit
4 new batteries.
2. Reinsert the batteries with the +/- poles
the right way round.
3. Clean the battery contacts with a dry
cloth.
Fit new batteries.
Check that the air tube is properly
connected to the monitor. Push the tube
connector firmly into the socket.
Contact your Omron retailer or distributor.
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron 705CP-2 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron 705CP-2 in the language / languages: English as an attachment in your email.

The manual is 0,41 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron 705CP-2

Omron 705CP-2 User Manual - Dutch - 24 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info